The PXIe-4150 features an Access LED and Voltage LED.
Access LED
The Access LED, located on the module front panel, indicates module power and access.
The following table lists the Access LED states.
| Status Indicator | Device State |
|---|---|
| (Off) | Not Powered |
| Green | Powered |
| Amber | Device is being accessed |
Why Is the Access LED Off When the Chassis Is On?
The LEDs may not light until the module has been configured in Hardware Configuration Utility or MAX. Before proceeding, verify that the PXIe-4150 appears in Hardware Configuration Utility or MAX.
If the Access LED fails to light after you power on the chassis, a problem may exist with the chassis power rails, a hardware module, or the LED.
- Disconnect any signals from the module front panel.
- Power off the chassis.
-
Remove the module from the chassis and inspect it for damage.
Notice Do not reinstall a damaged module.
- Install the module in a different, supported slot within the same PXI chassis.
-
Power on the chassis.
Note If you are using a PC with a device for PXI remote control system, power on the chassis before powering on the computer.
- Verify that the module appears in Hardware Configuration Utility or MAX.
- Reset the module in Hardware Configuration Utility or MAX and perform a self-test.
Voltage LED
The Voltage LED on the module front panel indicates the output channel state.
The following table lists the Voltage LED states.
| Status Indicator | Output Channel State |
|---|---|
| (Off) | The device output is disconnected from the voltage generation source through output disconnect relays. |
| Green | The device output is connected to the voltage generation source and <42.4 V DC is present. |
| Amber | The device output is connected to the voltage generation source and ≥42.4 V DC is present. High voltage may be generated by the device itself or an external device. |
| Red | The device has a fault or is in error due to the voltage generated or measured by the device. Refer to the driver software for possible sources. The device will not operate until the error is cleared and/or the device is reset. High voltage may be present on the device. |
| Blinking Green | The device output is in an inhibited state because the INHIBIT pin is
not shorted to ground. You must short the INHIBIT pin to ground to
restore normal operation. Note If ≥42.4 V DC is present, the LED
shines solid amber regardless of the INHIBIT pin
state. |
Related Information
- INHIBIT Pin
The INHIBIT input pin forces the PXIe-4150 output to an inhibited state if not shorted to ground. This inhibited state forces the PXIe-4150 to source 0 V, with a current limit of 1% of the configured device current range.