PXIe-4136 Specifications
- Updated2025-04-29
- 16 minute(s) read
PXIe-4136 Specifications
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Warranted unless otherwise noted.
Conditions
Specifications are valid under the following conditions unless otherwise noted.
- Ambient temperature[1]1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake). of 23 °C ± 5 ºC
- Calibration interval of 1 year
- 30 minutes warm-up time
- Self-calibration performed within the last 24 hours
- NI-DCPower Aperture Time property is set to 2 power-line cycles (PLC)
- Fans set to the highest setting if the PXI Express chassis has multiple fan speed settings
Cleaning Statement
PXIe-4136 Pinout
The following figure shows the terminals on the PXIe-4136 connector.
| Signal Name | Description |
|---|---|
| Output LO | LO force terminal connected to channel power stage (generates and/or dissipates power). Positive polarity is defined as voltage measured on HI > LO. |
| Sense LO | Voltage remote sense input terminals. Used to compensate for I*R voltage drops in cable leads, connectors, and switches. |
| Guard | Buffered output that follows the voltage of the HI force terminal. Used to drive shield conductors surrounding HI force and Sense HI conductors to minimize effects of leakage and capacitance on low level currents. |
| Output HI | HI force terminal connected to channel power stage (generates and/or dissipates power). Positive polarity is defined as voltage measured on HI > LO. |
| Sense HI | Voltage remote sense input terminals. Used to compensate for I*R voltage drops in cable leads, connectors, and switches. |
Device Capabilities
The following table and figure illustrate the voltage and the current source and sink ranges of the PXIe-4136.
| DC voltage ranges | DC current source and sink ranges |
|---|---|
|
DC sourcing power is limited to 20 W, regardless of output voltage.[3]3 Power limit defined by voltage measured between HI and LO terminals.
Voltage
Tcal is the internal device temperature recorded by the PXIe-4136 at the completion of the last self-calibration.
| Range | Resolution (noise limited) | Noise (0.1 Hz to 10 Hz, peak to peak), Typical | Accuracy (23 °C ± 5 °C) ± (% of voltage + offset) [4]4 Accuracy is specified for no load output configurations. Refer to Load Regulation and Remote Sense sections for additional accuracy derating and conditions. | Tempco ± (% of voltage + offset)/°C, 0 °C to 55 °C |
|---|---|---|---|---|
| Tcal ± 5 °C | ||||
| 600 mV | 1 μV | 4 μV | 0.020% + 100 μV | 0.0005% + 1 μV |
| 6 V | 10 μV | 12 μV | 0.020% + 640 μV | |
| 20 V | 100 μV | 40 μV | 0.022% + 2 mV | |
| 200 V | 1 mV | 400 μV | 0.025% + 20 mV |
Current
Tcal is the internal device temperature recorded by the PXIe-4136 at the completion of the last self-calibration
| Range | Resolution (noise limited) | Noise (0.1 Hz to 10 Hz, peak to peak), Typical | Accuracy (23 °C ± 5 °C) ± (% of current + offset) | Tempco ± (% of current + offset)/°C, 0 °C to 55 °C |
|---|---|---|---|---|
| Tcal ± 5 °C | ||||
| 1 μA | 1 pA | 8 pA | 0.03% + 200 pA | 0.0006% + 4 pA |
| 10 μA | 10 pA | 60 pA | 0.03% + 1.4 nA | 0.0006% + 22 pA |
| 100 μA | 100 pA | 400 pA | 0.03% + 12 nA | 0.0006% + 200 pA |
| 1 mA | 1 nA | 4 nA | 0.03% + 120 nA | 0.0006% + 2 nA |
| 10 mA | 10 nA | 40 nA | 0.03% + 1.2 μA | 0.0006% + 20 nA |
| 100 mA | 100 nA | 400 nA | 0.03% + 12 μA | 0.0006% + 200 nA |
| 1 A | 1 μA | 4 μA | 0.04% + 120 μA | 0.0006% + 2 μA |
Noise
Wideband source noise | <20 mV peak-to-peak in 20 V range, device configured for normal transient response, 10 Hz to 20 MHz, typical |
Sinking Power vs. Ambient Temperature Derating
The following figure illustrates sinking power derating as a function of ambient temperature for the PXIe-4136.
Overvoltage Protection
Accuracy[5]5 Overvoltage protection accuracy is valid with an ambient temperature of 23 °C ± 5 °C and with Tcal ±5 °C. Tcal is the internal device temperature recorded by the PXIe-4136 at the completion of the last self-calibration. (% of OVP limit + offset) | 0.1% + 200 mV, typical |
Temperature coefficient (% of OVP limit + offset)/°C | 0.01% + 3 mV/°C , typical |
Measurement location | Local sense |
Maximum OVP limit value | 210 V |
Minimum OVP limit value | 2 V |
Transient Response and Settling Time
Settling time is measured as the time to settle to within 0.1% of step amplitude, device configured for fast transient response.
Transient response | <70 μs to recover within 0.1% of voltage range after a load current change from 10% to 90% of range, device configured for fast transient response, typical | ||||||||||||
| |||||||||||||
The following figures illustrate the effect of the transient response setting on the step response of the PXIe-4136 for different loads.
Load Regulation
| |||||||
Current, device configured for local or remote sense | Load regulation effect included in current accuracy specifications, typical |
Expected Relay Life
Output Connected | ≥100 k cycles |
Measurement and Update Timing Characteristics
Pulse on time is measured from the start of the leading edge to the start of the trailing edge.
Pulse off time is measured from the start of the trailing edge to the start of a subsequent leading edge.
Available sample rates[6]6 When sourcing while measuring, both the Source Delay and Aperture Time affect the sampling rate. When taking a measure record, only the Aperture Time affects the sampling rate. | (1.8 MS/s)/N where N = 1, 2, 3, … 224, nominal | ||||||||||||
Sample rate accuracy | Equal to PXIe_CLK100 accuracy, nominal | ||||||||||||
Maximum measure rate to host | 1.8 MS/s per channel, continuous, nominal | ||||||||||||
| |||||||||||||
| |||||||||||||
| |||||||||||||
Remote Sense
Voltage accuracy | Add 3 ppm of voltage range per volt of HI lead drop plus 1 μV per volt of lead drop per ohm of corresponding sense lead resistance to voltage accuracy specifications |
Maximum sense lead resistance | 100 Ω |
Maximum lead drop per lead | 3 V, maximum 202 V between HI and LO terminals |
Safety Interlock
The safety interlock feature is designed to prevent users from coming in contact with hazardous voltage generated by the SMU in systems that implement protective barriers with controlled user access points.
| |||||||
| |||||||
Examples of Calculating Accuracy Specifications
Specifications listed in examples are for demonstration purposes only and do not necessarily reflect specifications for this device.
Tcal is the internal device temperature recorded by the PXIe-4136 at the completion of the last self-calibration.
Example 1: Calculating 5 °C Accuracy
Calculate the accuracy of 900 nA output in the 1 µA range under the following conditions:
| Ambient temperature | 28 °C |
| Internal device temperature | within Tcal ±5 °C |
| Self-calibration | within the last 24 hours |
Because the device internal temperature is within Tcal ±5 °C and the ambient temperature is within 23 °C ±5 °C, the appropriate accuracy specification is the following value:
0.03% + 200 pA
Calculate the accuracy using the following formula:
Therefore, the actual output is within 470 pA of 900 nA.
Example 2: Calculating Remote Sense Accuracy
Calculate the remote sense accuracy of 500 mV output in the 600 mV range. Assume the same conditions as in Example 1, with the following differences:
| HI path lead drop | 3 V |
| HI sense lead resistance | 2 Ω |
| LO path lead drop | 2.5 V |
| LO sense lead resistance | 1.5 Ω |
Because the device internal temperature is within Tcal ±5 °C and the ambient temperature is within 23 °C ±5 °C, the appropriate accuracy specification is the following value:
0.02% + 100 μV
Because the device is using remote sense, use the following remote sense accuracy specification:
Add 3 ppm of voltage range plus 11 μV per volt of HI lead drop plus 1 μV per volt of lead drop per Ω of corresponding sense lead resistance to voltage accuracy specifications.
Calculate the remote sense accuracy using the following formula:
Therefore, the actual output is within 248.2 µV of 500 mV.
Example 3: Calculating Accuracy with Temperature Coefficient
Calculate the accuracy of 900 nA output in the 1 µA range. Assume the same conditions as in Example 1, with the following differences:
| Ambient temperature | 15 °C |
Because the device internal temperature is within Tcal ±5 °C, the appropriate accuracy specification is the following value:
0.03% + 200 pA
Because the ambient temperature falls outside of 23 °C ±5 °C, use the following temperature coefficient per °C outside the 23 °C ±5 °C range:
0.0006% + 4 pA
Calculate the accuracy using the following formula:
Therefore, the actual output is within 498.2 pA of 900 nA.
Trigger Characteristics
Input triggers
Types | Start, Source, Sequence Advance, Measure, Pulse | ||||||
| |||||||
| |||||||
Output triggers (events)
Types | Source Complete, Sequence Iteration Complete, Sequence Engine Done, Measure Complete, Pulse Complete, Ready for Pulse | ||||||
| |||||||
Protection
Output channel protection
Overcurrent or overvoltage | Automatic shutdown, output disconnect relay opens |
Sink overload protection | Automatic shutdown, output disconnect relay opens |
Overtemperature | Automatic shutdown, output disconnect relay opens |
Safety interlock | Disable high voltage output, output disconnect relay opens |
Safety Voltage and Current
DC voltage | ±200 V | ||||||
| |||||||
Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.
DC current range | ±1 A |
Guard Output Characteristics
| |||||||
Calibration Interval
Recommended calibration interval | 1 year |
Power Requirement
PXI Express power requirement | 2.5 A from the 3.3 V rail and 2.7 A from the 12 V rail |
Physical
Dimensions | 3U, one-slot, PXI Express/CompactPCI Express module 2.0 cm × 13.0 cm × 21.6 cm (0.8 in. × 5.1 in. × 8.5 in.) |
Weight | 419 g (14.8 oz) |
Front panel connectors | 5.08 mm (8 position) |
Safety interlock connector | 3.55 mm (4 position) |
Environmental Guidelines
Environmental Characteristics
| Operating | 0 °C to 55 °C |
| Storage | -40 °C to 71 °C |
| Operating | 10% to 90%, noncondensing |
| Storage | 5% to 95%, noncondensing |
| Pollution degree | 2 |
| Maximum altitude | 2,000 m (800 mbar) (at 25 °C ambient temperature) |
| Operating vibration | 5 Hz to 500 Hz, 0.3 g RMS |
| Non-operating vibration | 5 Hz to 500 Hz, 2.4 g RMS |
| Operating shock | 30 g, half-sine, 11 ms pulse |
Safety Compliance Standards
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- AS/NZS CISPR 11: Group 1, Class A emissions
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
EU and UK Customers
电子信息产品污染控制管理办法(中国RoHS)
Product Certifications and Declarations
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake).
2 Voltage levels and limits > |40 V DC| require the safety interlock input to be closed.
3 Power limit defined by voltage measured between HI and LO terminals.
4 Accuracy is specified for no load output configurations. Refer to Load Regulation and Remote Sense sections for additional accuracy derating and conditions.
5 Overvoltage protection accuracy is valid with an ambient temperature of 23 °C ± 5 °C and with Tcal ±5 °C. Tcal is the internal device temperature recorded by the PXIe-4136 at the completion of the last self-calibration.
6 When sourcing while measuring, both the Source Delay and Aperture Time affect the sampling rate. When taking a measure record, only the Aperture Time affects the sampling rate.
7 As the source delay is adjusted or if advanced sequencing is used, maximum source rates vary. Timed output mode is enabled in Sequence Mode by setting Sequence Step Delta Time Enabled to True.
8 Shorter minimum on times for in-range pulses can be achieved using Sequence mode or Timed Output mode with Output Function set to Voltage or Current.
9 Pulses fall inside DC limits.
10 Pulse widths and logic levels are compliant with PXI Express Hardware Specification Revision 1.0 ECN 1.
11 Input triggers can be re-exported.