PXIe-4133 Device Specifications
- Updated2025-11-08
- 14 minute(s) read
PXIe-4133 Specifications
Definitions
Warranted describes the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristic describes values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical describes the performance met by a majority of models.
- Nominal describes an attribute that is based on design, conformance testing, or supplemental testing.
Values are Nominal unless otherwise noted.
Conditions
- Ambient temperature[1]1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake) or the specified operating temperature of your PXI chassis, whichever is lower.: 0 °C to 55 °C
- Chassis slot cooling capacity: ≥58 W[2]2 The PXIe-4133 must be installed in a chassis with slot cooling capacity ≥58 W.
- Calibration interval: 2 years
Device Capabilities
The PXIe-4133 is intended to test both forward and reverse bias characteristics of diodes. The following tables describe the basic capabilities in each mode of operation.
Current Ranges | 10 A, 2.5 A, 0.5 A, 0.1 A |
Voltage Range | 15 V |
Pulse Width | 2.5 µs to 655 µs (subject to current amplitude and duty cycle limits) |
Current Ranges | 100 µA, 10 µA, 1 µA |
Voltage Range | 100 V |
Pulse Width | ≤ 524 ms |
Current Range | 10 µA |
Voltage Range | 100 V |
Pulse Width | ≤ 524 ms |
Pulsed Current and Voltage Measurement Operations
| Current Range | Output Resolution |
Output Accuracy (% of Current + Offset) |
Measure Accuracy (% of Current + Offset) |
Measure Accuracy (% of Voltage + Offset) |
|||
|---|---|---|---|---|---|---|---|
| Typical | Maximum | Typical | Maximum | Typical | Maximum | ||
| 10 A | 176 µA | ±0.4% ±1 mA | ±0.7% ±3.5 mA | ±0.2% ±1.5 mA | ±0.35% ±5 mA | ±0.05% ±2 mV | ±0.1% ±3.5 mV |
| 2.5 A | 44 µA | ±0.4% ±0.4 mA | ±0.7% ±1 mA | ±0.2% ±1 mA | ±0.35% ±5 mA | ||
| 0.5 A | 8.8 µA | ±0.25% ±0.15 mA | ±0.4% ±0.3 mA | ±0.2% ±0.1 mA | ±0.35% ±0.25 mA | ||
| 0.1 A | 1.8 µA | ±0.15% ±0.03 mA | ±0.3% ±0.05 mA | ±0.15% ±0.02 mA | ±0.25% ±0.05 mA | ||
| Accuracy specifications are valid from 0 °C to +55 °C,
within 2 years of external calibration. Output accuracy is valid with output voltages of +0.5 V to +15 V. Voltage measurement range:0 V to +15 V, remotely sensed. Voltage measure accuracy is valid with sense voltages relative to
Force LO:
|
|||||||
Output current rise time | 0.8 µs, 10% to 90% of output |
Output current overshoot | 0.15% of final settled value |
Minimum pulse width | 2.5 µs |
Maximum pulse width | 655 µs (subject to current amplitude and duty cycle limits[3]3 For information about duty cycle and current amplitude limits, refer to Pulse Width and Duty Cycle Limitations in the PXIe-4133 User Manual.) |
Maximum pulse period[4]4 The pulse period includes the pulse width and dead time. | 20.971 ms |
Pulse width resolution | 10 ns |
Measure sampling rate (waveform mode) | 2 MS/s |
Measure aperture time (average point per pulse mode) | 0 ms to 524.29 ms |
Measure start time range (immediate pulse trigger[5]5 For an immediate trigger, specify none as the pulse trigger source.) | -20.971 ms to 20.971 ms (relative to pulse start) |
Measure start time range (external pulse trigger) | 0 ms to 20.971 ms (relative to pulse start) |
Measure start time resolution[6]6 Measure start time is the time of the first sample in a waveform or the start time of the measurement aperture (in average point per pulse mode). Measure start time is independently programmable for each pulse. | 10 ns |
Maximum pulse count per sequence | 10,000 |
Output protection | In case of an open circuit or overvoltage (Force HI to Force LO voltage >25 V) for more than 4 µs after the start of a pulse, the output disconnects and the pulse sequence immediately terminates. |
Repeatability Specifications
Non-linearity, temperature stability, and voltage measurement sensitivity errors are all included in the total accuracy specifications. They are included here to provide a measure of repeatability across variations in operating conditions.
| Current Range | Temperature Stability |
Voltage Measurement Sensitivity |
||
|---|---|---|---|---|
| Current Output | Current Measurement |
to Output Current (±0.25 mV Maximum over All Currents) |
to Common Mode Voltage | |
| 10 A | (±0.006% of output ± 60 µA) per °C | (±0.005% of reading ± 80 µA) per °C | ±0.025 mV per A | ±0.5 mV per volt across Force LO cable |
| 2.5 A | (±0.006% of output ± 15 µA) per °C | (±0.005% of reading ± 80 µA) per °C | ±0.025 mV per A | |
| 0.5 A | (±0.004% of reading ± 5 µA) per °C | (±0.005% of reading ± 5 µA) per °C | ±0.5 mV per A | |
| 0.1 A | (±0.002% of reading ± 1 µA) per °C | (±0.004% of reading ± 1 µA) per °C | ±2.5 mV per A | |
Temperature stability (voltage measurement, 15 V range) | (±0.0012% of reading ± 30 µV) per °C |
Non-linearity (current output) | ±0.006% of range |
Non-linearity (current measurement) | ±0.003% of range |
Non-linearity (voltage measurement) | ±0.001% of range |
Reverse Current Measurements
| Current Range | Output Voltage Range | Measure Accuracy (% of current + offset) | |||
|---|---|---|---|---|---|
| 0 V > Vout > -10 V | Any Output Voltage | ||||
| Typical | Maximum | Typical | Maximum | ||
| -100 µA | -0.1 V to -20 V | ±0.1% ±10 nA | ±0.15% ±30 nA | ±0.1% ±10 nA | ±0.15% ±32 nA |
| -10 µA | -0.1 V to -100 V | ±0.1% ±3.0 nA | ±0.2% ±10 nA | ±0.1% ±3.0 nA | ±0.2% ±28 nA |
| -1 µA | -0.1 V to -100 V | ±0.1% ±0.5 nA | ±0.3% ±1.5 nA | ±0.1% ±0.5 nA | ±0.3% ±3.3 nA |
| Accuracy specifications are valid from 0 to +55 °C, within 2 years of external calibration | |||||
Maximum pulse width | 524 ms |
Output protection | Outputs are protected against open circuits (measured current is within offset error specifications) and short circuits (outputs are current limited). |
| Voltage Range | Output Resolution | Output Accuracy (% of Voltage + Offset + Load Regulation) | Measure Accuracy (% of Voltage + Offset) | ||
|---|---|---|---|---|---|
| Typical | Maximum | Typical | Maximum | ||
| -100 V | 1.64 mV | ±0.07% ±10mV +0.5 mV/µA | ±0.15% ±20 mV +0.5 mV/µA | ±0.05% ±10 mV | ±0.1% ±30 mV |
| The reverse voltage output has a 500 Ω output impedance that causes the output voltage to become more positive (closer to zero) with larger load current. | |||||
Repeatability Specifications
Non-linearity, temperature stability, and voltage measurement sensitivity errors are all included in the total accuracy specifications. They are included here to provide a measure of repeatability across variations in operating conditions.
| Current Range | Current Measurement Temperature Stability |
Current Measurement Sensitivity to Output Voltage |
|---|---|---|
| -100 µA | (±0.003% of reading ± 0.2 nA) per °C | ±0.2 nA/V |
| -10 µA | (±0.003% of reading ± 0.1 nA) per °C | ±0.2 nA/V |
| -1 µA | (±0.005% of reading ± 0.02 nA) per °C | ±0.02 nA/V |
Temperature stability (voltage output) | (±0.0015% of reading ± 0.25 mV) per °C |
Temperature stability (voltage measurement) | (±0.0012% of reading ± 0.3 mV) per °C |
Non-linearity (voltage output) | ±0.003% of range |
Non-linearity (voltage measurement) | ±0.01% of range |
Non-linearity (current measurement) | ±0.01% of range |
Reverse Voltage Measurements
| Output Current Range | Output Current Resolution | Output Accuracy (% of Current + Offset) | Measure Voltage Range | Measure Accuracy (% of voltage + offset) | ||||
|---|---|---|---|---|---|---|---|---|
| 0 V > Vout > -60 V | Any Output Voltage | |||||||
| Typical | Maximum | Typical | Maximum | Typical | Maximum | |||
| ‑0.1 uA to ‑10 uA | 0.2 nA | ±0.1% ±10 nA | ±0.2% ±35 nA | ±0.1% ±15 nA | ±0.2% ±55 nA | 0 V to 100 V | ±0.05% ±10 mV | ±0.1% ±30 mV |
| Accuracy specifications are valid from 0 to +55 °C, within 2 years of external calibration | ||||||||
Maximum pulse width | 524 ms |
Output protection | Programmable output voltage limit (Vclamp). Valid values range from 0 V to -100 V. |
The reverse output bias current maintains accuracy for load voltages within the specified output clamp voltage. At load voltages beyond (more negative than) the specified clamp voltage, the output current bias current is reduced. The output bias current shuts off and reaches zero current typically between 0.5 V and 2.5 V beyond the programmed clamp voltage. The following figure depicts an example of varying output load voltage with the clamp voltage set to -50 V, for three different output bias current settings.
Repeatability Specifications
Non-linearity, temperature stability, and voltage measurement sensitivity errors are all included in the total accuracy specifications. They are included here to provide a measure of repeatability across variations in operating conditions.
Non-linearity (current output) | ±0.003% of range |
Non-linearity (voltage measurement) | ±0.01% of range |
Temperature stability (current output) | (±0.002% of output ± 0.3 nA) per °C |
Temperature stability (voltage measurement) | (±0.0012% of reading ± 0.3 mV) per °C |
Current output sensitivity to output voltage | ±0.5 nA/V |
Current output sensitivity to (Vout-Vclamp) difference | ±0.3 nA/V (Vout-Vclamp < 70 V) ±0.5 nA/V (Vout-Vclamp > 70 V) |
Contact Check Functions
The PXIe-4133 supports two methods of contact checking: the Low Voltage Contact Check function and the Contact Check function. The Low Voltage Contact Check more strictly limits the available output voltage and current. This may be important when probing potentially sensitive circuits or structures. The Contact Check function supports higher voltages and currents, which may be desirable when trying to forward bias a diode junction during the contact check.
Summary of the contact check methods:
Low Voltage Contact Check
- Outputs voltage up to 0.5 V between Force_HI and Force_LO
- Short circuit limited to 650 μA
- Measures current (0 to 100 μA range) out of Force_HI pin
- Measures voltage between Sense_HI and Sense_LO pins (remote sense, 4-wire mode)
Contact Check
- Outputs current up to 10 mA from Force_HI to Force_LO
- Voltage clamp programmable up to 5 V between Force_HI and Force_LO
- Measures current into Force_LO pin
- Measures voltage across Force pins (local sense, 2-wire mode) or Sense pins (remote sense, 4-wire mode)
With the Contact Check function, the PXIe-4133 pulses a test current in the same manner as the normal Pulse Current mode, but with the addition of an output voltage clamp. Current is driven out of the Force_HI pin and returns through and is measured at the Force_LO pin. You can read back the voltage either from the local sense (Force_HI and Force_LO, 2-wire mode) or from remote sense (Sense_HI and Sense_LO, 4-wire mode). The voltage clamp is placed between the Force_HI and Force_LO (across the current source) so selecting between local and remote sense has no impact on the current or the output voltage.
In the Contact Check mode, the current source has a 1 kOhm output impedance, and the voltage clamp has a 50 Ohm impedance. The following figure shows the resulting Current vs Voltage profile with the output current set to 10 mA and voltage clamps set for 1 V to 5 V.
The low voltage contact check function limits the possible voltage and current to much lower energy levels for validating contact with more sensitive circuits and structures. The Low Voltage contact check applies a small configurable bias voltage (up to 500 mV) between Force_HI and Force_LO. The current that flows out of the Force_HI pin is measured within a range of 0 to 100 μA, and is limited to about 650 μA in case of short circuits.
Low Voltage Contact Check also reads the voltage between Sense_HI and Sense_LO. Sense_HI and Sense_LO are only used to read voltage; the voltage on the Sense lines does not impact the Force current and voltage. Low Voltage Contact Check does not support Local Sense mode; the voltage between the Force pings is not read back directly.
In the Low Voltage Contact Check mode, the configurable output voltage has an impedance of about 120 Ohms. There is a 640 μA current limiter with an impedance of 5 kOhm. The following figure shows the resulting Current vs Voltage profile with the output voltage at settings between 50 mV and 500 mV.
Triggers
| Signal | Action | Source | Supported Delay |
|---|---|---|---|
| Start Trigger |
Starts the pulse sequence. |
None, PXI trigger lines (PXI_Trig0-7), PFI0, PFI1 | None |
| Pulse Trigger |
Generates the next pulse in the pulse sequence. |
None, PXI trigger lines (PXI_Trig0-7), PFI0, PFI1 | 0 s to 5 s |
| Abort Trigger | Aborts the pulsing within ≤7 µs upon receiving the falling edge. | None, PXI trigger lines (PXI_Trig0-7), PFI0, PFI1 | None |
| Signal | Action | Destination | Supported Delay | Minimum Pulse Width | Notes |
|---|---|---|---|---|---|
| Ready for Start Trigger Event | Asserts when the PXIe-4133 is ready to start the pulse sequence. | None, PXI trigger lines (PXI_Trig0-7), PFI0, PFI1 | None | 400 ns | The start trigger must be configured with a source other than none. |
| Ready for Pulse Trigger Event | Asserts when the PXIe-4133 is ready to generate the next pulse in the sequence. | None, PXI trigger lines (PXI_Trig0-7), PFI0, PFI1 | None | 400 ns | The start pulse must be configured with a source other than none. |
| Pulse Event | Asserts with each pulse. | None, PXI trigger lines (PXI_Trig0-7), PFI0, PFI1 | ±20.971 ms | 400 ns | – |
Data clock resolution | 10 ns |
Timing reference | PXIe_CLK100 |
Front Panel Connectors
| Number of connectors | 4 total connectors, 2 connectors per channel (1 force connection and 1 sense connection) |
|
| Connector type | TRB (triaxial BNC), 3-lug, 120 degree keying | |
| Terminal Assignments | Center Pin | HI |
| Middle Ring | LO | |
| Outer Shell | Ground | |
| Number of connectors | 2 (PFI<0, 1>) |
| Connector type | SMB |
| Direction | Bidirectional |
| Inputs (Triggers) | Start Trigger, Pulse Trigger, Abort Trigger |
| Outputs (Events) | Ready for Start Trigger Event, Ready for Pulse Trigger Event, Pulse Event |
Rated Voltages
The PXIe-4133 is intended to measure passive devices, such as diodes, in configurations where voltages are either generated by the PXIe-4133 directly or by currents output from the PXIe-4133 which are driven across a device under test.
The maximum voltages and currents that the PXIe-4133 can output to a VCSEL device being tested are:
| Specification | Value |
|---|---|
| Forward Bias | +10 A and +15 Vpk |
| Reverse Bias | -100 µA and -100 Vpk |
The maximum voltage levels into the trigger inputs:
| Specification | Value |
|---|---|
| PFI-to-Ground | 0 V to +5 V |
Measurement Category
This product is rated for Measurement Category I.
Measurement Category I
Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.
Environmental Guidelines
Environmental Characteristics
| Temperature | Operating | 0 °C to 55 °C |
| Storage | -40 °C to 71 °C | |
| Humidity | Operating | 10% RH to 90% RH, noncondensing |
| Storage | 5% RH to 95% RH, noncondensing | |
| Pollution Degree | 2 | |
| Maximum altitude | 2000 m | |
| Shock and vibration | Operating vibration | 5 Hz to 500 Hz, 0.3 g RMS |
| Non-operating vibration | 5 Hz to 500 Hz, 2.4 g RMS | |
| Operating shock | 30 g, half-sine, 11 ms pulse | |
Calibration Interval
| Recommended calibration interval | 2 years |
Power Requirements
| Specification | Value |
|---|---|
| Minimum chassis slot power | 58 W |
Power Specifications
| Specification | Value |
|---|---|
| +3.3 V load, maximum | 1.5 A |
| +12 V load, maximum | 4.5 A (peak), 3.2 A (average) |
| Maximum rated power | 43 W |
Physical Characteristics
Dimensions and Weight
| Dimensions |
3U, one-slot, PXI Express/CompactPCI Express module 2.0 cm × 13.0 cm × 21.6 cm (0.8 in. × 5.1 in. × 8.5 in.) |
| Weight | 521 g |
1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake) or the specified operating temperature of your PXI chassis, whichever is lower.
2 The PXIe-4133 must be installed in a chassis with slot cooling capacity ≥58 W.
3 For information about duty cycle and current amplitude limits, refer to Pulse Width and Duty Cycle Limitations in the PXIe-4133 User Manual.
4 The pulse period includes the pulse width and dead time.
5 For an immediate trigger, specify none as the pulse trigger source.
6 Measure start time is the time of the first sample in a waveform or the start time of the measurement aperture (in average point per pulse mode). Measure start time is independently programmable for each pulse.