Definitions

Warranted Specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical—describes the performance met by a majority of models.
  • Nominal—describes an attribute that is based on design, conformance testing, or supplemental testing.
  • Values are Typical unless otherwise noted.

    Conditions

    Specifications are valid under the following conditions unless otherwise noted.

    • Indoor use only
    • The product is operated within the Environmental limits described in this document, including:
      • Altitude and pollution degree as specified in Environmental Characteristics
      • Ambient temperature and relative humidity as specified in Operating Environment (per-slot power profiles apply)

    PXIe-1081 Pinouts

    Use the pinouts to connect to terminals on the PXIe-1081 backplane.

    System Controller Slot XP1 Connector Pinout

    Pins Signals
    A GND
    B 12V
    C 12V
    D GND
    E 5V
    F 3.3V
    G GND

    System Controller Slot XP2 Connector Pinout

    Pin A B ab C D cd E F ef
    1 2PETp1 2PETn1 GND 2PERp1 2PERn1 GND 2PETp2 2PETn2 GND
    2 2PETp3 2PETn3 GND 2PERp3 2PERn3 GND 2PERp2 2PERn2 GND
    3 2PETp4 2PETn4 GND 2PERp4 2PERn4 GND 2PETp5 2PETn5 GND
    4 2PETp6 2PETn6 GND 2PERp6 2PERn6 GND 2PERp5 2PERn5 GND
    5 2PETp7 2PETn7 GND 2PERp7 2PERn7 GND 2PETp8 2PETn8 GND
    6 2PETp9 2PETn9 GND 2PERp9 2PERn9 GND 2PERp8 2PERn8 GND
    7 2PETp10 2PETn10 GND 2PERp10 2PERn10 GND 2PETp11 2PETn11 GND
    8 2PETp12 2PETn12 GND 2PERp12 2PERn12 GND 2PERp11 2PERn11 GND
    9 2PETp13 2PETn13 GND 2PERp13 2PERn13 GND 2PETp14 2PETn14 GND
    10 2PETp15 2PETn15 GND 2PERp15 2PERn15 GND 2PERp14 2PERn14 GND

    System Controller Slot XP3 Connector Pinout

    Pin A B ab C D cd E F ef
    1 RSV RSV GND RSV RSV GND RSV RSV GND
    2 RSV RSV GND PWR_OK PS_ON# GND LINKCAP PWRBTN# GND
    3 SMBDAT SMBCLK GND RSVD RSVD GND RSVD RSVD GND
    4 RSV PERST# GND 2RefClk+ 2RefClk- GND 1RefClk+ 1RefClk- GND
    5 1PETp0 1PETn0 GND 1PERp0 1PERn0 GND 1PETp1 1PETn1 GND
    6 1PETp2 1PETn2 GND 1PERp2 1PERn2 GND 1PERp1 1PERn1 GND
    7 1PETp3 1PETn3 GND 1PERp3 1PERn3 GND 1PETp4 1PETn4 GND
    8 1PETp5 1PETn5 GND 1PERp5 1PERn5 GND 1PERp4 1PERn4 GND
    9 1PETp6 1PETn6 GND 1PERp6 1PERn6 GND 1PETp7 1PETn7 GND
    10 2PETp0 2PETn0 GND 2PERp0 2PERn0 GND 1PERp7 1PERn7 GND

    System Controller Slot XP4 Connector Pinout

    Pin Z A B C D E F
    1 GND GA4 GA3 GA2 GA1 GA0 GND
    2 GND 5Vaux GND SYSEN# WAKE# ALERT# GND
    3 GND RSV RSV RSV RSV RSV GND
    4 GND RSV RSV RSV RSV RSV GND
    5 GND PXI_TRIG3 PXI_TRIG4 PXI_TRIG5 GND PXI_TRIG6 GND
    6 GND PXI_TRIG2 GND RSV PXI_STAR PXI_CLK10 GND
    7 GND PXI_TRIG1 PXI_TRIG0 RSV GND PXI_TRIG7 GND
    8 GND RSV GND RSV RSV PXI_LBR6 GND

    Hybrid Slot P1 Connector Pinout

    Pin Z A B C D E F
    25 GND 5V REQ64# ENUM# 3.3V 5V GND
    24 GND AD[1] 5V V(I/O) AD[0] ACK64# GND
    23 GND 3.3V AD[4] AD[3] 5V AD[2] GND
    22 GND AD[7] GND 3.3V AD[6] AD[5] GND
    21 GND 3.3V AD[9] AD[8] M66EN C/BE[0]# GND
    20 GND AD[12] GND V(I/O) AD[11] AD[10] GND
    19 GND 3.3V AD[15] AD[14] GND AD[13] GND
    18 GND SERR# GND 3.3V PAR C/BE[1]# GND
    17 GND 3.3V IPMB_SCL IPMB_SDA GND PERR# GND
    16 GND DEVSEL# GND V(I/O) STOP# LOCK# GND
    15 GND 3.3V FRAME# IRDY# BD_SEL# TRDY# GND
    12 to 14 Key Area
    11 GND AD[18] AD[17] AD[16] GND C/BE[2]# GND
    10 GND AD[21] GND 3.3V AD[20] AD[19] GND
    9 GND C/BE[3]# IDSEL AD[23] GND AD[22] GND
    8 GND AD[26] GND V(I/O) AD[25] AD[24] GND
    7 GND AD[30] AD[29] AD[28] GND AD[27] GND
    6 GND REQ# GND 3.3V CLK AD[31] GND
    5 GND BRSVP1A5 BRSVP1B5 RST# GND GNT# GND
    4 GND IPMB_PWR HEALTHY# # V(I/O) INTP INTS GND
    3 GND INTA# INTB# INTC# 5V INTD# GND
    2 GND TCK 5V TMS TDO TDI GND
    1 GND 5V -12V TRST# +12V 5V GND

    Hybrid Slot XP3 Connector Pinout

    Pin A B ab C D cd E F ef
    1

    PXIe_

    CLK100+

    PXIe_

    CLK100-
    GND

    PXIe_

    SYNC100+

    PXIe_

    SYNC100-
    GND

    PXIe_

    DSTARC+

    PXIe_

    DSTARC-
    GND
    2 PRSNT# PWREN# GND

    PXIe_

    DSTARB+

    PXIe_

    DSTARB-
    GND

    PXIe_

    DSTARA+

    PXIe_

    DSTARA-
    GND
    3 SMBDAT SMBCLK GND RSV RSV GND RSV RSV GND
    4 MPWRGD* PERST# GND RSV RSV GND 1RefClk+ 1RefClk- GND
    5 1PETp0 1PETn0 GND 1PERp0 1PERn0 GND 1PETp1 1PETn1 GND
    6 1PETp2 1PETn2 GND 1PERp2 1PERn2 GND 1PERp1 1PERn1 GND
    7 1PETp3 1PETn3 GND 1PERp3 1PERn3 GND 1PETp4 1PETn4 GND
    8 1PETp5 1PETn5 GND 1PERp5 1PERn5 GND 1PERp4 1PERn4 GND
    9 1PETp6 1PETn6 GND 1PERp6 1PERn6 GND 1PETp7 1PETn7 GND
    10 RSV RSV GND RSV RSV GND 1PERp7 1PERn7 GND

    Hybrid Slot XP4 Connector Pinout

    Pin Z A B C D E F
    1 GND GA4 GA3 GA2 GA1 GA0 GND
    2 GND 5Vaux GND SYSEN# WAKE# ALERT# GND
    3 GND 12V 12V GND GND GND GND
    4 GND GND GND 3.3V 3.3V 3.3V GND
    5 GND

    PXI_

    TRIG3

    PXI_

    TRIG4

    PXI_

    TRIG5
    GND

    PXI_

    TRIG6
    GND
    6 GND

    PXI_

    TRIG2
    GND ATNLED

    PXI_

    STAR

    PXI_

    CLK10
    GND
    7 GND

    PXI_

    TRIG1

    PXI_

    TRIG0
    ATNSW# GND

    PXI_

    TRIG7
    GND
    8 GND RSV GND RSV

    PXI_

    LBL6

    PXI_

    LBR6
    GND

    Physical Characteristics

    Table 1. Physical Characteristics
    Dimensions 3U, 21-slot-wide, PXI Express chassis 177.1 mm × 445.5 mm × 463.6 mm (6.97 in. × 17.54 in. × 18.25 in.)

    For more information, visit ni.com/dimensions and search by module number.

    Weight 13.5 kg (29.8 lb)
    Chassis materials Sheet Aluminum (5052-H32, 5754-H22), Extruded Aluminum (6063-T5, 6060-T6), Plate Aluminum (6063-T5, 6061-T6), Cold Rolled Steel, Cold Rolled Stainless Steel, Sheet Copper (C110), Santoprene, Urethane Foam, PC-ABS, Nylon, Polycarbonate, Delrin, Polyethylene, Polyamide (FR-106), Neodymium Magnet
    Chassis finish Conductive Clear Iridite on Aluminum, Electroplated Nickel on Cold Rolled Steel, Electroplated Zinc on Cold Rolled Steel, Electroplated Nickel on Copper

    Environmental Guidelines

    Notice Failure to follow the mounting instructions in the product documentation can cause temperature derating.
    Notice This product is intended for use in indoor applications only.
    Notice Cover all empty slots using filler panels.

    Environmental Characteristics

    Table 2. Operating Environment
    Ambient temperature range when all modules require ≤38 W cooling capacity per slot 0 °C to 50 °C
    Ambient temperature range when any module requires >38 W cooling capacity per slot (up to 58 W) 0 °C to 40 °C
    Relative humidity range 10% to 90%, noncondensing
    Table 3. Storage Environment
    Ambient temperature range –40 °C to 71 °C
    Relative humidity range 5% to 95%, noncondensing
    Table 4. Pollution Degree
    Pollution degree 2
    Table 5. Maximum Altitude
    Maximum altitude 2,000 m (6,560 ft.), 800 mbar (at 25 °C ambient, high fan mode)
    Table 6. Shock and Vibration
    Operating shock 30 g peak, half-sine, 11 ms pulse
    Operational random vibration 5 to 500 Hz, 0.3 grms
    Non-operating vibration 5 to 500 Hz, 2.4 grms
    Table 7. Acoustic Emissions
    Profile Fan Mode Sound Pressure Level (at Operator Position) Sound Power Level
    38 W Auto (≤30 °C) 34.4 dBA 47.5 dBA
    38 W High 55.0 dBA 66.3 dBA
    58 W Auto (≤30 °C) 52.1 dBA 63.7 dBA
    58 W High 66.3 dBA 77.4 dBA
    Note This product meets the requirements of the environmental standards for electrical equipment for measurement, control, and laboratory use.
    Note Acoustic emissions testing was conducted per ISO 7779. The results meet the compliance criteria outlined in MIL-PRF-28800F.
    Caution The protection provided by the PXIe-1081 can be impaired if it is used in a manner not described in this document.

    Electrical

    The following section provides information about the PXIe-1081 AC input and DC output.

    AC Input

    Table 8. AC Input Characteristics
    Input rating 100 to 240 VAC, 50/60 Hz, 10 to 5 A, 100 to 120 VAC, 400 Hz, 10 A
    Operating voltage range 90 to 264 VAC
    Nominal input frequency 50 Hz/60 Hz/400 Hz
    Operating frequency range 47 to 440 Hz
    Efficiency 85%, typical
    Over-current protection Internal fuse in line
    Main power disconnect The AC power cable provides main power disconnect. Do not position the equipment so that it is difficult to disconnect the power cord. The front-panel power switch causes the internal chassis power supply to provide DC power to the PXI Express backplane.
    Caution Disconnect power cord to completely remove power.
    Note The operating range is guaranteed by design, with the exception that 400 Hz operation is only supported from 100 to 120 VAC.

    DC Output

    Table 9. DC Output Characteristics by Voltage Rail
    Voltage Rail Maximum Current Load Regulation Maximum Ripple and Noise (20 MHz BW)
    +5V_AUX 4.1 A ±5% 50 mVpp
    +12 V 54.1 A ±5% 100 mVpp
    +5 V 51.1 A ±5% 50 mVpp
    +3.3 V 60 A ±5% 50 mVpp
    -12 V 4.25 A ±5% 50 mVpp
    Note Maximum usable power for the PXIe-1081 is 650 W.
    Table 10. Backplane Slot Current Capacity
    Slot +5 V V (I/O) +3.3 V +12 V -12 V 5 VAUX
    PXI Express System Controller Slot 9 A - 9 A 20 A - 2 A
    Hybrid Peripheral Slot with PXI-5 Peripheral - - 6 A 6 A - 1 A
    Hybrid Peripheral Slot with PXI-1 Peripheral 6 A 5 A 6 A 1 A 1 A -
    Note Total System Controller Slot current must not exceed 40 A.
    Note PCI V (I/O) pins in Hybrid Peripheral Slots are connected to +5 V.
    Note The maximum power dissipated in the System Controller Slot must not exceed 140 W.
    Note The maximum power dissipated in a peripheral slot must not exceed 58 W. Refer to Operating Environment for ambient temperature considerations at 58 W.
    Table 11. Power Supply Protection and Serviceability
    Feature Description
    Overcurrent protection All outputs protected from short circuit and overload with automatic recovery.
    Overvoltage protection + 12 V, +5 V, and +3.3 V clamped at 20 to 30% above nominal output voltage.
    Serviceability Fan replacement supported.

    Backplane

    Table 12. Backplane Characteristics
    Size 3U-sized; one system slot (with three system expansion slots) and 17 peripheral slots. Compliant with IEEE 1101.10 mechanical packaging. PXI Express Specification compliant. Accepts both PXI Express and CompactPCI (PICMG 2.0 R 3.0) 3U modules.
    Backplane bare-board material UL 94 V-0 Recognized
    Backplane connectors Conforms to IEC 917 and IEC 1076-4-101, UL 94 V-0 rated

    Chassis Cooling

    Table 13. Module Cooling
    Cooling method Forced air circulation (positive pressurization) through three 210 CFM fans with High/Auto speed selector
    Airflow direction Bottom of module to top of module
    Intake Bottom rear of chassis
    Exhaust Top of chassis
    Slot cooling capacity 58 W
    Table 14. Secondary Cooling
    Cooling method Forced air circulation (positive pressurization) through one 70 CFM fan
    Intake Right side of chassis
    Exhaust Left side of chassis
    Table 15. Power Supply Cooling
    Cooling method Forced air circulation through two integrated fans
    Intake Top rear of chassis
    Exhaust Top of chassis
    Table 16. Minimum Cooling Clearances
    Location Clearance
    Above 44.45 mm (1.75 in.)
    Rear 101.60 mm (4.00 in.)
    Sides 44.45 mm (1.75 in.)

    10 MHz System Reference Clock: PXI_CLK10

    Table 17. 10 MHz System Reference Clock: PXI_CLK10
    Maximum slot-to-slot skew 250 ps
    Accuracy ±25 ppm maximmum (guaranteed over the operating temperature range)
    Maximum jitter 5 ps RMS phase-jitter (10 Hz–1 MHz range)
    Duty-factor 45% to 55%
    Unloaded signal swing 3.3 V ±0.3 V
    Note For other specifications, refer to the latest version of the PXI Express Hardware Specification.

    100 MHz System Reference Clock: PXIe_CLK100 and PXIe_SYNC100

    Table 18. 100 MHz System Reference Clock: PXIe_CLK100 and PXIe_SYNC100
    Maximum slot-to-slot skew 100 ps
    Accuracy ±25 ppm maximum (guaranteed over the operating temperature range)
    Maximum jitter
    • 10 Hz to 12 kHz range:3 ps RMS phase-jitter
    • 12 kHz to 20 MHz range: 2 ps RMS phase-jitter
    Duty-factor for PXIe_CLK100 45% to 55%
    Absolute differential voltage (When terminated with a 50 Ω load to 1.30 V or Thévenin equivalent) 400 mV to 1,000 mV
    Note For other specifications, refer to the latest version of the PXI Express Hardware Specification.