PXI-2542 Specifications
- Updated2025-05-22
- 7 minute(s) read
PXI-2542 Specifications
PXI-2542 Specifications
Topology | Quad 2 × 1 multiplexers |
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
- Measured specifications describe the measured performance of a representative model.
Specifications are Warranted unless otherwise noted.
Conditions
Specifications are valid under the following conditions unless otherwise noted.
- Chassis is powered on
- Calibration adjustment cycle maintained
- 50 Ω termination connected to unused I/O front panel connectors
- All voltages are specified in DC, ACpk, or a combination unless otherwise specified.
Cleaning Statement
Clean devices and terminal blocks by brushing off light dust with a soft, nonmetallic brush. Remove other contaminants with a soft, lint-free, dampened cloth. Do not use detergent or chemical solvents. The unit must be completely dry and free from contaminants before returning to service.
PXI-2542 Pinout
| Signal | Description |
|---|---|
| CHxA and CHxB | Interchangeable signal connections |
| COMx | Routing destination for the corresponding channel |
Input Characteristics
Minimum input frequency | 10 MHz |
Characteristic impedance (Z0) | 50 Ω nominal |
Coupling | AC |
Maximum safe DC input voltage | ±8V |
Maximum Safe Continuous RF Power
Chassis power ON | +30 dBm |
Chassis power OFF | +20 dBm |
RF Performance Characteristics
Values in parentheses are typical.
| |||||||||
Insertion loss thermal coefficient | α = 1,112 ppm/°C | ||||||||
Use the following equation to calculate the insertion loss at a given temperature:
| where | IL is insertion loss in dB |
| T is the temperature at which the property is being measured in °C | |
| T 0 is the reference temperature in °C | |
| α represents insertion loss temperature coefficient in ppm/°C |
| Frequency | Flatness / 200 MHz | Flatness / 1GHz |
|---|---|---|
| 10 MHz to 5 GHz | <0.15 dB | <0.50 dB |
| 5 GHz to 6 GHz | <0.18 dB | <0.62 dB |
| 6 GHz to 6.6 GHz | <0.63 dB | <1.25 dB |
| |||||||||||
| |||||||||||
| |||||||||||
| |||||||||
Typical channel-to-channel skew | <3 ps | ||||||
Typical propagation delay | 680 ps | ||||||
| |||||||
Refer to the following figures for typical insertion loss, typical VSWR, and typical isolation, and typical bank-to-bank isolation respectively.
Linearity
Second-order harmonic distortion (Input IP2 (IIP2))
Typical IP2 (input) | >+91 dBm |
Third-order intermodulation distortion (Input IP3 (IIP3))
Typical IP3 (input) | >+61 dBm |
Dynamic Characteristics
Maximum switch operate time | 121 µs |
Trigger Characteristics
| |||||||
| |||||||
Physical Characteristics
Switch type | FET | ||||||||||||||||||
Front panel I/O connectors | 12 SMA jacks, female | ||||||||||||||||||
| |||||||||||||||||||
Dimensions (L × W × H) | 3U, one slot, PXI/cPCI module, PXI Express compatible 21.6 cm × 2.0 cm × 13.0 cm (8.5 in. × 0.8 in. × 5.1 in.) | ||||||||||||||||||
Weight | 553 g (19.5 oz) | ||||||||||||||||||
Environment
| Maximum altitude | 2,000 m (800 mbar) (at 25 °C ambient temperature) |
|---|---|
| Pollution Degree | 2 |
Indoor use only.
Operating Environment
| Ambient temperature range, operating | 0 °C to 55 °C |
|---|---|
| Relative humidity range, operating | 10% to 90%, noncondensing |
Storage Environment
| Ambient temperature, storage | -40 °C to 71 °C |
|---|---|
| Relative humidity range, storage | 5% to 95%, noncondensing |
Shock and Vibration
| Operating shock | 30 g peak, half-sine, 11 ms pulse |
|---|---|
| Random vibration |
|
Compliance and Certifications
Safety Compliance Standards
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1
Electromagnetic Compatibility
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- EN 55022 (CISPR 22): Class A emissions
- EN 55024 (CISPR 24): Immunity
- AS/NZS CISPR 11: Group 1, Class A emissions
- AS/NZS CISPR 22: Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
Product Certifications and Declarations
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.