Initializes an EEPROM-specific calibration step.

Supported Devices: PXIe-5601/5603/5605/5606, PXIe-5693/5694/5698


icon

Inputs/Outputs

  • civrn.png instrument handle

    instrument handle identifies your calibration instrument session. instrument handle is obtained from the niRFSA Init Ext Cal VI.

  • ci32.png calibration step

    calibration step specifies the calibration step to initialize.

    IF Attenuation Calibration (1600)

    Initializes the IF Attenuation Calibration step. This step is not supported for the PXIe-5693.

    IF Response Calibration (1601)

    Initializes the IF Response Calibration step. This step is not supported for the PXIe-5603/5605/5606, PXIe-5693/5698.

    Ref Level Calibration (1602)

    Initializes the Ref Level Calibration step. This step is not supported on the PXIe-5694.

    LO Export Calibration (1603)

    Initializes the LO Export Calibration step. This step calibrates the output power of each LO to be within specification. This step is not supported on the PXIe-5601 or the PXIe-5693/5694/5698.

    Gain Reference Calibration (1604)

    Initializes the Gain Reference Calibration step. This step calibrates the calibration tone amplitude across supported calibration tone frequencies. This step is not supported on the PXIe-5601/5603/5605/5606, PXIe-5694.

  • cerrcodeclst.png error in

    error in describes error conditions that occur before this node runs. This input provides standard error in functionality.

  • iivrn.png instrument handle out

    instrument handle out passes a reference to your instrument session to the next VI. instrument handle is obtained from the niRFSA Initialize VI.

  • ierrcodeclst.png error out

    error out contains error information. This output provides standard error out functionality.