Returns the date and time of the last successful self-calibration.
The time returned is 24-hour local time and the date is returned as integer values. For example, if the device was calibrated at 2:30 PM on December 31, 2010, this VI returns 14 for the hour output, 30 for the minute output, 12 for the month output, 31 for the day output, and 2010 for the year output.
Supported Devices: PXI-5600, PXIe-5601/5603/5605/5606 (external digitizer mode), PXIe-5644/5645/5646, PXI-5661, PXIe-5663/5663E/5665/5667/5668, PXIe-5820/5830/5831/5832/5840/5841/5842/5860

Inputs/Outputs
instrument handle
—
instrument handle identifies your calibration instrument session. instrument handle is obtained from the niRFSA Initialize VI, niRFSA Initialize With Options VI, or niRFSA Init Ext Cal VI.
self cal step
—
self cal step specifies the self-calibration step to query for the last successful self-calibration date and time data.
error in
—
error in describes error conditions that occur before this node runs. This input provides standard error in functionality.
year
—
year returns the year of the last self-calibration.
instrument handle out
—
instrument handle out passes a reference to your instrument session to the next VI. instrument handle is obtained from the niRFSA Initialize VI or the niRFSA Initialize With Options VI and identifies a particular instrument session.
month
—
month returns the month of the last self-calibration.
day
—
day returns the day of the last self-calibration.
minute
—
minute returns the minute of the last self-calibration.
error out
—
error out contains error information. This output provides standard error out functionality.
hour
—
hour returns the hour of the last self-calibration. |