NI-SCOPE Properties

Triggering:Trigger Window:Low Level Property

  • Updated2023-02-21
  • 1 minute(s) read

Short Name: Trigger Window Low Level

Property of niScope

Pass the lower voltage threshold you want the digitizer to use for window triggering.

The digitizer triggers when the trigger signal enters or leaves the window you specify with this property and the Trigger Window High Level property.

The values of the Vertical Range property and the Vertical Offset property determine the valid range for this property on the channel you specify with the Trigger Source property.

The value you pass for this parameter must meet the following conditions:

Low Trigger Level <= Vertical Range/2 + Vertical Offset

Low Trigger Level >= (-Vertical Range/2) + Vertical Offset

Low Trigger Level < High Trigger Level

Related topics:

Remarks

The following table lists the characteristics of this property.

Datatype fD
PermissionsRead/Write
High-level VIs niScope Configure Trigger Window (Analog Window Ref Trigger)
Channel-based No
ResettableNo

Log in to get a better experience