Configures common properties for runt triggering.
A runt trigger occurs when both the leading edge and trailing edge of a pulse cross only one of two trigger thresholds you specify and with a polarity you specify, where the polarity is relative to the threshold crossed.

Inputs/Outputs
trigger coupling
—
trigger coupling specifies how to couple the trigger signal. Refer to the Trigger Coupling property for defined values.
polarity
—
polarity specifies the polarity of the runt pulses, relative to the runt threshold the pulses cross, that trigger the oscilloscope for runt triggering. Refer to the Runt Polarity property for defined values.
instrument handle
—
instrument handle identifies a particular instrument session.
trigger source (Channel 0)
—
trigger source passes the source you want the digitizer to monitor for a trigger. Default Value: Channel 0 Refer to the Trigger Source property for defined values.
low threshold
—
low threshold specifies, in volts, the lower of two thresholds that bound the vertical range to examine for runt pulses. Refer to the Runt Low Threshold property for more information.
high threshold
—
high threshold specifies, in volts, the higher of two thresholds that bound the vertical range to examine for runt pulses. Refer to the Runt High Threshold property for more information.
error in
—
error in describes error conditions that occur before this node runs. This input provides standard error in functionality.
trigger holdoff
—
trigger holdoff specifies the length of time the digitizer waits after detecting a trigger before enabling NI-SCOPE to detect another trigger. Default Value: 0.0 s
trigger delay
—
trigger delay specifies how long the digitizer waits after it receives the trigger to start acquiring data. Default Value: 0.0 s
instrument handle out
—
instrument handle out has the same value as the instrument handle.
error out
—
error out contains error information. This output provides standard error out functionality. |
trigger coupling
—
polarity
—
instrument handle
—
trigger source (Channel 0)
—
low threshold
—
error in
—
instrument handle out
—