Triggering:Trigger Runt:Runt Time Condition
- Updated2025-10-21
- 1 minute(s) read
Specifies whether runt triggers are time qualified, and if so, how the oscilloscope triggers in relation to the duration range bounded by the Runt Time Low Limit and Runt Time High Limit properties.
Default Value: None
Remarks
The following table lists the characteristics of this property.
| Short Name | Runt Time Condition |
| Data type | ![]() |
| Permissions | Read/Write |
| High-level VIs | N/A |
| Channel-based | No |
| Resettable | No |
| None | 0 | Time qualification is disabled. The oscilloscope triggers on runt pulses based solely on the voltage level of the pulses. |
| Within | 1 | The oscilloscope triggers on pulses that, in addition to meeting runt voltage criteria, have a duration within the range bounded by Runt Time Low Limit and Runt Time High Limit. |
| Outside | 2 | The oscilloscope triggers on pulses that, in addition to meeting runt voltage criteria, have a duration not within the range bounded by Runt Time Low Limit and Runt Time High Limit. |