niScope Cal Self Calibrate VI
- Updated2025-10-21
- 2 minute(s) read
Self-calibrates most NI digitizers, including all SMC-based devices.
For SMC-based devices, if the self-calibration is performed successfully in a regular session, the calibration constants are immediately stored in the self-calibration area of the EEPROM. If the self-calibration is performed in an external calibration session, the calibration constants take effect immediately for the duration of the session. However, they are not stored in the EEPROM until you call niScope Cal End with action set to Store Calibration and no errors occur.

Inputs/Outputs
instrument handle
—
instrument handle identifies a particular instrument session.
channels
—
channels specifies the channel(s) to calibrate. For more information, refer to Channel String Syntax.
Option
—
Option allows you to self-calibrate all channels or restore the external calibration. Defined Values Self-Calibrate All Channels (Default) Restore External Calibration
error in
—
error in describes error conditions that occur before this node runs. This input provides standard error in functionality.
instrument handle out
—
instrument handle out has the same value as the instrument handle.
error out
—
error out contains error information. This output provides standard error out functionality. |