Closes an external calibration session. You must call this VI each time you call niScope Cal Start, even if an error occurs during calibration.


icon

Inputs/Outputs

  • civrn.png instrument handle

    instrument handle is the instrument handle that you obtain from niScope Cal Start. The handle identifies a particular instrument session.

  • ci32.png action

    action either stores the calibration constants or aborts the calibration and discards any calibration results.

    Store Calibration (0) Stores the new calibration constants in the EEPROM. For most oscilloscopes, the current system date and the incremented calibration count are also stored. For SMC-based oscilloscopes, the current system date and onboard temperature are also stored.
    Abort Calibration (1) Closes the session and discards any new calibration constants. Some devices may write to the EEPROM during calibration, in which case this action restores the EEPROM to its original state.
  • cerrcodeclst.png error in

    error in describes error conditions that occur before this node runs. This input provides standard error in functionality.

  • ierrcodeclst.png error out

    error out contains error information. This output provides standard error out functionality.