NI-9320 Specifications

These specifications apply to the NI-9320.

Revision History

Version Date changed Description
379260A-01 November 2025 Initial release.

Looking For Something Else?

For information not found in the specifications for your product, such as operating instructions, browse Related Information.

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.

Specifications are Typical unless otherwise noted.

Conditions

Specifications are valid for the range -40 °C to 70 °C unless otherwise noted. All voltages are relative to the AI- signal on each channel unless otherwise noted.

NI-9320 Push-In Spring Terminal Pinout



Table 1. Signal Descriptions
Signal Description
AI+ Positive analog input signal connection
AI- Negative analog input signal connection
COM Common reference connection to isolated ground

NI-9320 with DSUB Pinout



Table 2. Signal Descriptions
Signal Description
AI+ Positive analog input signal connection
AI- Negative analog input signal connection
COM Common reference connection to isolated ground

Input Characteristics

Number of channels 16 analog input channels
ADC resolution 16 bits
Type of ADC Successive approximation register (SAR)
No missing codes 16 bits guaranteed
Table 3. Input Voltage Ranges
Measurement voltage (AI+ to AI-) ±10.5 V (max)
Maximum common mode voltage ±10.5 V (max)
Overvoltage protection ±30 V
Conversion time 5 μs
Sample rate 200 kS/s
Table 4. Accuracy
Measurement Conditions Within 2 Years of Calibration Within 10 Years of Calibration
Percent of Reading (Gain Error) Percent of Range[1]1 Range equals ±10.5 V. For example, maximum offset error within 2 years of calibration equals 10.5 V * +/-0.022% = +/- 2.3 mV. (Offset Error) Percent of Reading (Gain Error) Percent of Range[1] (Offset Error) 
Maximum (-40 °C to 70 °C) ±0.093% ±0.022% ±0.144% ±0.030%
Typical (23 °C ±5 °C) ±0.015% ±0.003% ±0.026% ±0.004%
Note NI recommends a calibration interval of 2 years and only provides calibration services referencing the 2-year specification limits. If you choose to calibrate less often, the 10-year column provides predicted performance over this extended interval. Choose an appropriate interval based on your application requirements. Longer calibration intervals are more likely to result in As-Found calibration failures when the device is sent back for calibration and compared against the 2-year specification limits.
Table 5. Stability Drift
Gain drift 2.5 ppm/°C (Typical)
Offset drift 6.5 μV/°C (Typical)
Common-mode rejection ratio (CMRR) (fin = 60 Hz) 70 dB
-3 dB bandwidth[2]2 Up to 3.5 Vrms sine wave. > 200 kHz
Input impedance > 1 GΩ
Input noise 0.5 LSBrms
Crosstalk -105 dB
Table 6. Settling Time (to 2 LSBs)
10 V step 12 μs
20 V step 15 μs

NI-9320 with Push-In Spring Terminal (Black/Orange Connector) Safety Voltages

Connect only voltages that are within the following limits.

Channel-to-channel

±10.5 V maximum

Channel-to-COM

±10.5 V maximum

Isolation

Channel-to-channel

None

Channel-to-COM

None

Channel-to-earth ground

Continuous

250 V RMS, Measurement Category II

Withstand up to 4000 m

4250 V DC, verified by a 5 s dielectric withstand test

NI-9320 with DSUB Safety Voltages

Connect only voltages that are within the following limits.

Channel-to-channel

±10.5 V maximum

Channel-to-COM

±10.5 V maximum

Isolation

Channel-to-channel

None

Channel-to-COM

None

Channel-to-earth ground

Continuous

60 V DC, Measurement Category I

Withstand up to 3000 m

1000 V RMS, verified by a 5 s dielectric withstand test

Withstand up to 5000 m

860 V RMS, verified by a 5 s dielectric withstand test

Measurement Category

Measurement Category I

Caution Do not connect the NI-9320 with DSUB to signals or use for measurements within Measurement Categories II, III, or IV.
Attention Ne pas connecter le NI-9320 with DSUB à des signaux dans les catégories de mesure II, III ou IV et ne pas l'utiliser pour effectuer des mesures dans ces catégories.
Warning Do not connect the NI-9320 with DSUB to signals or use for measurements within Measurement Categories II, III, or IV, or for measurements on MAINs circuits or on circuits derived from Overvoltage Category II, III, or IV which may have transient overvoltages above what the product can withstand. The product must not be connected to circuits that have a maximum voltage above the continuous working voltage, relative to earth or to other channels, or this could damage and defeat the insulation. The product can only withstand transients up to the transient overvoltage rating without breakdown or damage to the insulation. An analysis of the working voltages, loop impedances, temporary overvoltages, and transient overvoltages in the system must be conducted prior to making measurements.
Mise en garde Ne pas connecter le NI-9320 with DSUB à des signaux dans les catégories de mesure II, III ou IV et ne pas l'utiliser pour des mesures dans ces catégories, ou des mesures sur secteur ou sur des circuits dérivés de surtensions de catégorie II, III ou IV pouvant présenter des surtensions transitoires supérieures à ce que le produit peut supporter. Le produit ne doit pas être raccordé à des circuits ayant une tension maximale supérieure à la tension de fonctionnement continu, par rapport à la terre ou à d'autres voies, sous peine d'endommager et de compromettre l'isolation. Le produit peut tomber en panne et son isolation risque d'être endommagée si les tensions transitoires dépassent la surtension transitoire nominale. Une analyse des tensions de fonctionnement, des impédances de boucle, des surtensions temporaires et des surtensions transitoires dans le système doit être effectuée avant de procéder à des mesures.

Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.

Note Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are for other circuits not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.

Measurement Category II

Caution Do not connect the NI-9320 with spring terminal to signals or use for measurements within Measurement Categories III or IV.
Attention Ne pas connecter le NI-9320 with spring terminal à des signaux dans les catégories de mesure III ou IV et ne pas l'utiliser pour effectuer des mesures dans ces catégories.

Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.

Environmental Characteristics

Temperature

Operating

-40 °C to 70 °C

Storage

-40 °C to 85 °C

Humidity

Operating

10% RH to 90% RH, noncondensing

Storage

5% RH to 95% RH, noncondensing

Ingress protection

IP40

Pollution Degree

2

Maximum altitude

NI-9320 with push-in spring terminal (black/orange connector)

4,000 m

NI-9320 with DSUB

5,000 m

Shock and Vibration
Operating vibration

Random

5 g RMS, 10 Hz to 500 Hz

Sinusoidal

5 g, 10 Hz to 500 Hz

Operating shock

30 g, 11 ms half sine; 50 g, 3 ms half sine; 18 shocks at 6 orientations

To meet these shock and vibration specifications, you must panel mount the system.

Power Requirements

Power consumption from chassis (full-scale input, 200 kS/s)

Active mode

1 W maximum

Sleep mode

1 mW maximum

Thermal dissipation (at 70 °C)

Active mode

1.4 W maximum

Sleep mode

600 mW maximum

Physical Characteristics

Weight

NI-9320 with push-in spring terminal (black/orange connector)

165 g (5.8 oz)

NI-9320 with DSUB

150 g (5.3 oz)

NI-9320 with Push-In Spring Terminal (Black/Orange Connector)

Spring terminal wiring

Gauge

0.14 mm2 to 1.5 mm2 (26 AWG to 16 AWG) copper conductor wire

Wire strip length

10 mm (0.394 in.) of insulation stripped from the end

Temperature rating

90 °C minimum

Wires per spring-terminal

One wire per spring-terminal; two wires per spring terminal using a 2-wire ferrule

Ferrules

Single ferrule, uninsulated

0.14 mm2 to 1.5 mm2 (26 AWG to 16 AWG) 10 mm barrel length

Single ferrule, insulated

0.14 mm2 to 1.0 mm2 (26 AWG to 18 AWG) 12 mm barrel length

Two-wire ferrule, insulated

2 x 0.34 mm2 (2 x 22 AWG) 12 mm barrel length

Connector securement

Securement type

Screw flanges provided

Torque for screw flanges

0.2 N · m (1.80 lb · in.)

Calibration

You can obtain the calibration certificate and information about calibration services for the NI-9320 at ni.com/calibration.
Recommended calibration interval 2 years

1 Range equals ±10.5 V. For example, maximum offset error within 2 years of calibration equals 10.5 V * +/-0.022% = +/- 2.3 mV.

2 Up to 3.5 Vrms sine wave.