NI-9320 Specifications
- Updated2025-11-20
- 7 minute(s) read
NI-9320 Specifications
These specifications apply to the NI-9320.
Revision History
| Version | Date changed | Description |
|---|---|---|
| 379260A-01 | November 2025 | Initial release. |
Looking For Something Else?
For information not found in the specifications for your product, such as operating instructions, browse Related Information.
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Typical unless otherwise noted.
Conditions
Specifications are valid for the range -40 °C to 70 °C unless otherwise noted. All voltages are relative to the AI- signal on each channel unless otherwise noted.
NI-9320 Push-In Spring Terminal Pinout
| Signal | Description |
|---|---|
| AI+ | Positive analog input signal connection |
| AI- | Negative analog input signal connection |
| COM | Common reference connection to isolated ground |
NI-9320 with DSUB Pinout
| Signal | Description |
|---|---|
| AI+ | Positive analog input signal connection |
| AI- | Negative analog input signal connection |
| COM | Common reference connection to isolated ground |
Input Characteristics
| Number of channels | 16 analog input channels |
| ADC resolution | 16 bits |
| Type of ADC | Successive approximation register (SAR) |
| No missing codes | 16 bits guaranteed |
| Measurement voltage (AI+ to AI-) | ±10.5 V (max) |
| Maximum common mode voltage | ±10.5 V (max) |
| Overvoltage protection | ±30 V |
| Conversion time | 5 μs |
| Sample rate | 200 kS/s |
| Measurement Conditions | Within 2 Years of Calibration | Within 10 Years of Calibration | ||
|---|---|---|---|---|
| Percent of Reading (Gain Error) | Percent of Range[1]1 Range equals ±10.5 V. For example, maximum offset error within 2 years of calibration equals 10.5 V * +/-0.022% = +/- 2.3 mV. (Offset Error) | Percent of Reading (Gain Error) | Percent of Range[1] (Offset Error) | |
| Maximum (-40 °C to 70 °C) | ±0.093% | ±0.022% | ±0.144% | ±0.030% |
| Typical (23 °C ±5 °C) | ±0.015% | ±0.003% | ±0.026% | ±0.004% |
| Gain drift | 2.5 ppm/°C (Typical) |
| Offset drift | 6.5 μV/°C (Typical) |
| Common-mode rejection ratio (CMRR) (fin = 60 Hz) | 70 dB |
| -3 dB bandwidth[2]2 Up to 3.5 Vrms sine wave. | > 200 kHz |
| Input impedance | > 1 GΩ |
| Input noise | 0.5 LSBrms |
| Crosstalk | -105 dB |
| 10 V step | 12 μs |
| 20 V step | 15 μs |
NI-9320 with Push-In Spring Terminal (Black/Orange Connector) Safety Voltages
Connect only voltages that are within the following limits.
Channel-to-channel | ±10.5 V maximum | ||||||
Channel-to-COM | ±10.5 V maximum | ||||||
| |||||||
| |||||||
NI-9320 with DSUB Safety Voltages
Connect only voltages that are within the following limits.
Channel-to-channel | ±10.5 V maximum | ||||||||
Channel-to-COM | ±10.5 V maximum | ||||||||
| |||||||||
| |||||||||
Measurement Category
Measurement Category I
Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.
Measurement Category II
Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.
Environmental Characteristics
| |||||||||||||
| |||||||||||||
Ingress protection | IP40 | ||||||||||||
Pollution Degree | 2 | ||||||||||||
| |||||||||||||
| |||||||||||||
To meet these shock and vibration specifications, you must panel mount the system.
Power Requirements
| |||||||
| |||||||
Physical Characteristics
Weight
NI-9320 with push-in spring terminal (black/orange connector) | 165 g (5.8 oz) |
NI-9320 with DSUB | 150 g (5.3 oz) |
NI-9320 with Push-In Spring Terminal (Black/Orange Connector)
| |||||||||||||||||||||||||||||
Calibration
| Recommended calibration interval | 2 years |