NI-9263 Specifications

These specifications apply to the NI-9263.

Revision History

Version Date changed Description
378838B-01 August 2025 Added the product pinout and clarified the weight of the spring terminal variant.
378838A-01 June 2022 Initial release.

Looking For Something Else?

For information not found in the specifications for your product, such as operating instructions, browse Related Information.

Connector Types

The NI-9263 has more than one connector type: NI-9263 with screw terminal and NI-9263 with spring terminal. Unless the connector type is specified, NI-9263 refers to both connector types.

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.

Specifications are Typical unless otherwise noted.

Conditions

Specifications are valid under the following conditions unless otherwise noted.

  • -40 °C to 70 °C
  • All voltages are relative to COM

NI-9263 Pinout

Figure 1. NI-9263 Connector Pinouts

NI-9263 Connector Pinouts

Note To create a secure connection when connecting more than one wire to a single terminal on the NI-9263, use 2-wire ferrules.
Table 1. Signal Descriptions
Signal Name Description
AO Analog output signal connection
COM Common reference connection to isolated ground
NC No connection

Physical Characteristics

Table 2. Physical Characteristics
Dimensions

Visit ni.com/dimensions and search by module number.

Weight
  • NI-9263 with screw terminal: 150 g (5.3 oz)
  • NI-9263 with spring terminal: 139 g (4.9 oz)
Table 3. Screw-Terminal Wiring
Gauge 0.2 mm2 to 2.5 mm2 (26 AWG to 14 AWG) copper conductor wire
Wire strip length 13 mm (0.51 in.) of insulation stripped from the end
Temperature rating 90 °C, minimum
Torque for screw terminals 0.5 N · m to 0.6 N · m (4.4 lb · in. to 5.3 lb · in.)
Wires per screw terminal One wire per screw terminal; two wires per screw terminal using a 2-wire ferrule
Ferrules 0.25 mm2 to 2.5 mm2
Table 4. Spring-Terminal Wiring
Gauge 0.2 mm2 to 2.5 mm2 (26 AWG to 14 AWG) copper conductor wire
Wire strip length 10 mm (0.39 in.) of insulation stripped from the end
Temperature rating 90 °C, minimum
Torque for spring terminals 0.5 N · m to 0.6 N · m (4.4 lb · in. to 5.3 lb · in.)
Wires per spring terminal One wire per spring terminal; two wires per spring terminal using a 2-wire ferrule
Ferrules 0.25 mm2 to 2.5 mm2
Table 5. Connector Securement
Securement type Screw flanges provided
Torque for screw flanges 0.2 N · m (1.80 lb · in.)

Environmental Guidelines

Notice Failure to follow the mounting instructions in the product documentation can cause temperature derating.
Notice This model is intended for use in indoor applications only.

Environmental Characteristics

Table 6. Temperature
Characteristic Specification
Operating -40 °C to 70 °C
Storage -40 °C to 85 °C
Table 7. Humidity
Characteristic Specification
Operating 10% RH to 90 % RH, noncondensing
Storage 5% RH to 95 % RH, noncondensing
Table 8. Ingress Protection
Characteristic Specification
Ingress protection IP40
Table 9. Pollution Degree
Characteristic Specification
Pollution degree 2
Table 10. Maximum Altitude
Characteristic Specification
Maximum altitude
  • NI-9263 with screw terminal: 2,000 m
  • NI-9263 with spring terminal: 2,000 m
Table 11. Shock and Vibration
Operating vibration - Random 5 g RMS, 10 Hz to 500 Hz
Operating vibration - Sinusoidal 5 g, 10 Hz to 500 Hz
Operating shock 30 g, 11 ms half sine; 50 g, 3 ms half sine; 18 shocks at 6 orientations

To meet these shock and vibration specifications, you must panel mount the system.

Output Characteristics

Table 12. Output Characteristics
Number of channels 4 analog output channels
DAC resolution 16 bit
DAC type String
Power-on output state Channels off
Startup voltage[1]1 When the module powers on, a glitch occurs for 20 μs peaking at -1.5 V. 0 V
Power-down voltage[2]2 The power-down voltage peaks at 1.8 V before exponentially discharging to 0 V in 100 μs. To reduce the peak voltage, add a 10 kΩ load. 0 V
Current drive ±1 mA per channel maximum
Output impedance 2 Ω
Slew rate 4 V/μs
Crosstalk 76 dB
Capacitive drive 1,500 pF minimum
Monotonicity 16 bits
DNL ±1 LSB maximum
INL (endpoint) ±12 LSB maximum
MTBF 1,732,619 hours at 25 °C; Bellcore Issue 2, Method 1, Case 3, Limited Part Stress Method
Table 13. Output Voltage Range
Nominal ±10 V
Minimum ±10.4 V
Typical ±10.7 V
Maximum ±11 V
Table 14. Accuracy
Measurement Conditions Temperature Range Percent of Reading (Gain Error) Percent of Range[3]3 Range equals ±10.7 V (Offset Error)
Calibrated Maximum (-40 °C to 70 °C) 0.35% 0.75%
Calibrated Typical (25 °C, ±5 °C) 0.35% 0.75%
Uncalibrated [4]4 Uncalibrated accuracy refers to the accuracy achieved when acquiring in raw or unscaled modes where the calibration constants stored in the module are not applied to the data. Maximum (-40 °C to 70 °C) 2.2% 1.7%
Uncalibrated Typical (25 °C, ±5 °C) 0.3% 0.25%
Table 15. Stability
Gain drift 11 ppm/°C
Offset drift 110 μV/°C
Table 16. Protection
Overvoltage ±30 V
Short-circuit Indefinitely
Table 17. Update Time
Number of Channels Update Time for All Other Chassis Update Time for cRIO-9151 R Series Expansion Chassis
1 3 μs minimum 3.5 μs minimum
2 5 μs minimum 6.5 μs minimum
3 7.5 μs minimum 9 μs minimum
4 9.5 μs minimum 12 μs minimum
Table 18. Noise
Updating at 100 kS/s 600 μV RMS
Not updating 260 μV RMS
Table 19. Settling Time (100 pF load, to 1 LSB)
Full-scale step 20 μs
1 V step 13 μs
0.1 V step 10 μs

Power Requirements

Table 20. Power Consumption from Chassis
Active mode (at -40 °C) 500 mW maximum
Sleep mode 25 μW maximum
Table 21. Thermal Dissipation (at 70 °C)
Active mode 750 mW maximum
Sleep mode 25 μW maximum

Safety Voltages

Connect only voltages that are within the following limits:

Table 22. Channel to Channel
Channel-to-channel None
Table 23. Channel to Earth Ground
Continuous 250 V RMS, Measurement Category II
Withstand 2,300 V RMS, verified by a 5 s dielectric withstand test

Measurement Category II

Caution Do not connect the product to signals or use for measurements within Measurement Categories III or IV.
Attention Ne pas connecter le produit à des signaux dans les catégories de mesure III ou IV et ne pas l'utiliser pour effectuer des mesures dans ces catégories.

Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.

Calibration

You can obtain the calibration certificate and information about calibration services for the NI-9263 at ni.com/calibration.
Recommended calibration interval 1 year

1 When the module powers on, a glitch occurs for 20 μs peaking at -1.5 V.

2 The power-down voltage peaks at 1.8 V before exponentially discharging to 0 V in 100 μs. To reduce the peak voltage, add a 10 kΩ load.

3 Range equals ±10.7 V

4 Uncalibrated accuracy refers to the accuracy achieved when acquiring in raw or unscaled modes where the calibration constants stored in the module are not applied to the data.