Oscilloscopes
- Updated2026-01-29
- 11 minute(s) read
Oscilloscope Setup Actions
You can configure the setup of an oscilloscope with the Auto setup or the Default setup option.
To change setup mode, click the arrow next to the button that displays the currently selected setup mode.
Settings Changed by Auto Setup
Setting an oscilloscope to Auto setup changes the following settings to the following values:
| General Settings | Value |
|---|---|
| Acquisition Mode | Normal |
| Reference Clock | Internal |
| Vertical Settings | Value |
|---|---|
| Vertical Coupling | Unchanged by Auto setup |
| Vertical Bandwidth | Full |
| Vertical Range | Changed by Auto setup |
| Vertical Offset | 0 V |
| Probe Attenuation | Unchanged by Auto setup |
| Input Impedance | Unchanged by Auto setup |
| Horizontal Settings | Value |
|---|---|
| Sample Rate | Changed by Auto setup |
| Min Record Length | Changed by Auto setup |
| Enforce Realtime | True |
| Number of Records | 1 |
| Triggering Settings | Value |
|---|---|
| Trigger Type | Edge if signal present, otherwise Immediate |
| Trigger Channel | Lowest numbered channel with a signal present |
| Trigger Slope | Positive |
| Trigger Coupling | DC |
| Reference Position | 50% |
| Trigger Level | 50% of signal on trigger channel |
| Trigger Delay | 0 |
| Trigger Holdoff | 0 |
| Trigger Output | None |
Set the Sample Rate and the Record Length Manually
By default, InstrumentStudio optimizes the record length or the sample rate of an oscilloscope when taking measurements.
Configure the Record Length
Manual Override Mode
To configure larger records, use Manual Override mode. Manual Override mode allows for the configuration of both the sample rate and the record length of an oscilloscope.
Complete the following steps to enable Manual Override mode:
-
Click the
icon for the instrument to open the Instrument
Settings window.
-
On the Acquisition tab, click the Manual
Override mode toggle.
The toggle is in the On position.
- Enter the desired values for Sample Rate and Record Length.
Acquisition Status
You can determine the current state of an acquisition by looking at the acquisition status icon. The acquisition status icon is in the Horizontal & Acq. header of an oscilloscope panel.
Interpolation Methods
You can set the interpolation method of an oscilloscope acquisition from the Instrument Settings window. Click the dropdown menu for the Interpolation method setting in the Acquisition tab of the Instrument Settings window.
You can set interpolation to one of the following settings:
- The Random Interleaved Sampling (RIS) sampling method is set for the oscilloscope.
- The acquisition uses the peak detect sample mode.
- The waveform has enough data and does not need interpolation.
Sampling Methods
Depending on the device, you can select different sampling methods from the Instrument Settings window. Click the dropdown menu for the Sampling method setting in the Acquisition tab of the Instrument Settings window. You can set sampling to one of the following methods:
Sample Modes
Add FFT Channels and Markers
You can use FFT channels with an oscilloscope to measure the amplitude and the frequency of a sample. You can also fetch more detailed measurements and search for peaks with the help of markers.
Create FFT Channels
- From a large oscilloscope panel, click the FFT button in the Add Channels section. A frequency graph appears.
- Select a Source for the FFT channel. You can select any single channel that already exists.
- Select a Window function for the FFT channel.
- (Optional) Configure additional settings for
the FFT channel by clicking the
icon next to the FFT channel. For more information, see Search for and Compute Peaks.
Create Markers
A marker fetches the amplitude (y-location) of a particular sample at the specified frequency (x-location) of a particular sample.
Complete the following steps to create markers:
- Select the Markers dropdown menu in the Frequency Graph header.
- Select Markers: On to enable markers. The Markers table appears below the frequency graph.
- From the Markers table,
select the marker that you want to configure. By default, only Marker 0
is enabled. Note You can enable markers by selecting a marker from the Markers table. Then, change the marker mode from Off to Normal. You can enable up to 12 markers.
- Select the FFT channel that you want to measure with the selected marker.
- Enter the frequency where you want to place the
marker. The amplitude of the signal at the frequency of the marker is displayed in the
Level section of the marker toolbar. Note You can view and edit the frequency of all enabled markers in the Markers table. The Markers table also displays the current level, mode, and source channel of the markers.
Identify and Compute Peaks
To use the peak search functions to locate peaks, you must first enable at least one FFT
channel and one marker. The controls for the peak search functions (
) are on the left side of the Frequency tab.
- The highest peak
- The next highest peak
- A peak that is directly adjacent to the marker (both left or right)
Compute Peaks Using Peak Threshold and Peak Excursion
You can configure peak threshold and peak excursion settings by selecting the
icon in the Frequency tab. The
icon is directly to the right of the peak search functions (
).
Peak threshold defines the minimum amplitude level that a sample must rise above to be considered a peak. You can set a peak threshold value without setting a peak excursion value. Without peak excursion, InstrumentStudio considers a peak eligible for every amplitude measurement that is above the peak threshold value.
Peak excursion defines the minimum amplitude variation that is required in a signal to be considered a peak. Peak excursion is always specified in relation to a peak threshold value.
- The signal rises above the threshold level
- The rise of the signal above the peak threshold matches or exceeds the peak excursion value
- The fall of the signal from above the peak threshold matches or exceeds the peak excursion value
- The signal falls below the threshold level

| Peak | Peak Eligibility | Peak Characteristics |
|---|---|---|
| Peak 1 |
|
|
| Peak 2 |
|
|
| Peak 3 |
|
|
| Peak 4 |
|
|
| Peak 5 | ||
| Peak 6 |
|
|
| Peak 7 |
|
FFT Averaging Modes
You can perform averaged measurements on an FFT channel to improve measurement accuracy or to help compensate for a low signal-to-noise ratio.
To configure FFT averaging modes, click the
icon for the FFT channel, which opens the Software Channel
Settings window.
- Configuring a measurement system
- Applying limit testing to a frequency spectrum
- Applying upper limit testing to a frequency spectrum
Measure Power Spectral Density
Power spectral density measures the power content compared to the frequency of a signal. You can view the power spectral density of a signal when you analyze data in the frequency domain. Power spectral density helps you determine which frequency ranges have strong or weak variations in power.
- Add an oscilloscope to the large panel.
-
Create an FFT
channel by selecting the FFT button in the Add
Channels section of the large panel.
A frequency chart opens below the time chart of the large panel.
-
Configure the FFT Axis settings in the following way:
- In the header of the frequency chart, select the Chart Options button.
- In the Y-axis section, set Units to V/√(Hz) or dBm/Hz.
- In the X-axis section, set Scaling to Logarithmic.
-
Run the panel if it does not already run.
InstrumentStudio plots the power spectral density data on the Frequency chart.