GPIB-140B Specifications

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.

Specifications are Typical unless otherwise noted.

Conditions

Specifications are valid for 0 °C to 55 °C unless otherwise noted.

Cleaning Statement

Notice If you need to clean the GPIB-140B, wipe it with a dry towel. Make sure that the hardware is completely dry and free from contaminants before returning it to service.

System Configuration

Distance per extension

Up to 1 km

Loading per extension

Up to 13 additional devices (28 total devices in the extension system, including the extenders)

Multiple extension

Permitted in any pattern of star or linear pattern

Device Characteristics

Transmission interface unit

Optical transmitter and receiver (HFBR1414, HFBR2416, or equivalent) with ST-style optical cable connectors

GPIB interface load

Two standard loads, AC and DC

Performance Characteristics

Maximum transfer rate

Buffered mode, non-HS488

>1.1 MBytes/s

HS488 handshake

>2.8 MBytes/s

Unbuffered mode

>200 kBytes/s

Functionality

Transparent GPIB operation except for latched parallel polls

Interlocked IEEE 488 handshake

Maintained across the extension in unbuffered mode

IEEE 488 capability identification codes

SH1

Complete Source Handshake

AH1

Complete Acceptor Handshake

T5, TE5

Complete Talker

L3, LE3

Complete Listener

SR1

Complete Service Request

RL1

Complete Remote Local

PP1, 2

Complete Parallel Poll

DC1

Complete Device Clear

DT1

Complete Device Trigger

C1-5

Complete Controller

E2

Tri-state GPIB driver

HS488 capability identification codes

SHE

HS-488 Source Handshake

AHE

HS-488 Acceptor Handshake

Operational Characteristics

Architecture

Point-to-point (not multi-drop) transmission

Operating modes

Buffered or unbuffered (interlocked) mode

HS488 modes

Enabled HS488 or disabled HS488 mode

Parallel Poll Response modes

Immediate Parallel Poll Response mode or latched Parallel Poll Response mode

Power Requirements

Input voltage range

9 V DC to 15 V DC

Maximum power consumption

3.0 W (250 mA @ 12 V)

12 V DC Power Supply (NI PN 723595-02; shipped with the GPIB-140B)

Input voltage range

100 V AC to 240 V AC, 47 Hz to 63 Hz

GPIB-140B Pinout

Use the pinout to connect to terminals on the GPIB-140B.

Figure 1. GPIB-140B Connector Pinout


Table 1. Signal Descriptions
Signal Terminal Description
DIO1 1 Data Input/Output Bit.
DIO2 2 Data Input/Output Bit.
DIO3 3 Data Input/Output Bit.
DIO4 4 Data Input/Output Bit.
EOI 5 End-Or-Identify.
DAV 6 Data Valid.
NRFD 7 Not Ready For Data.
NDAC 8 Not Data Accepted.
IFC 9 Interface Clear.
SRQ 10 Service Request.
ATN 11 Attention.
SHIELD 12 Shield.
DIO5 13 Data Input/Output Bit.
DIO6 14 Data Input/Output Bit.
DIO7 15 Data Input/Output Bit.
DIO8 16 Data Input/Output Bit.
REN 17 Remote Enable.
GND 18 Ground—Wire twisted with DAV.
GND 19 Ground—Wire twisted with NRFD.
GND 20 Ground—Wire twisted with NDAC.
GND 21 Ground—Wire twisted with IFC.
GND 22 Ground—Wire twisted with SRQ.
GND 23 Ground—Wire twisted with ATN.
SIGNAL GROUND 24 Logic Ground.

Physical Characteristics

Dimensions and Weight

Overall case size (dimensions)

93.67 × 133.88 × 28.80 mm (3.69 × 5.27 × 1.13 in.)

Enclosure material

Aluminum

Weight

323 g (11.4 oz)

Field Wiring Specifications

GPIB cable

Type X2 shielded

Transmission cable

3.0 × 6.5 mm cable diameter

62.5/125 micron core/clad with NA = 0.275

850 nm operating wavelength​

3.0 dB/km attenuation Duplex style, terminated with ST-style connectors; multi-mode

Environmental Guidelines

Notice To achieve the full temperature range of the GPIB-140B, allow at least 1" of cooling clearance around the product during use. If it is not possible to follow this guideline, reduce the maximum temperature rating of the product by 5 °C.
Notice This product is intended for use in indoor applications only.

Environmental Characteristics

Table 2. Temperature
Operating0 °C to 55 °C
Storage-20 °C to 70 °C
Table 3. Humidity
Operating10% RH to 90% RH, noncondensing
Storage5% RH to 95% RH, noncondensing
Table 4. Pollution Degree
Pollution degree2
Table 5. Maximum Altitude
Maximum altitude2,000 m (800 mbar), at 25 °C ambient temperature
Table 6. Shock and Vibration
Operating vibration 5 Hz to 500 Hz, 0.3 g RMS
Non-operating vibration 5 Hz to 500 Hz, 2.4 g RMS
Operating shock30 g, half-sine, 11 ms pulse

Environmental Management

NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

EU and UK Customers

  • Waste Electrical and Electronic Equipment (WEEE)—At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.
  • 电子信息产品污染控制管理办法(中国RoHS)

  • 中国RoHSNI符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于NI中国RoHS合规性信息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)
  • Product Certifications and Declarations

    Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.

    NI Services

    Visit ni.com/support to find support resources including documentation, downloads, and troubleshooting and application development self-help such as tutorials and examples.

    Visit ni.com/services to learn about NI service offerings such as calibration options, repair, and replacement.

    Visit ni.com/register to register your NI product. Product registration facilitates technical support and ensures that you receive important information updates from NI.

    NI corporate headquarters is located at 11500 N Mopac Expwy, Austin, TX, 78759-3504, USA.