FD-11634 Specifications
- Updated2025-06-30
- 16 minute(s) read
FD-11634 Specifications
FD-11634 Specifications
These specifications apply to the FD-11634.
Revision History
| Version | Date changed | Description |
|---|---|---|
| 377651C-01 | June 2025 | Added pinout. |
| 377651B-01 | June 2019 | Updated butterworth filter table and graph. |
| 377651A-01 | February 2019 | Initial release. |
Looking For Something Else?
For information not found in the specifications for your product, such as operating instructions, browse Related Information.
Conditions
Specifications are typical and valid at -40 °C to 85 °C unless otherwise noted.
FD-11634 Front Panel
Use the front panel to locate the connectors, LEDs, and mounting holes on the FD-11634.
- Power IN Connector
- Power LED
- Ethernet Port 0 and LED
- Bank 1 Input Connectors 0 through 3 and LEDs
- Bank 2 Input Connectors 0 through 3 and LEDs
- Mounting Holes
- Power OUT Connector
- STATUS LED
- Ethernet Port 1 and LED
- SYNC Logo
FD-11634 Signal Input Connectors Pinout
The FD-11634 features eight 5-pin A-coded M12 connectors: Bank 1 connectors 0 through 3 and Bank 2 connectors 0 through 3. The following figure shows the pinout of a signal input connector.
| Pin Number | Wire Color* | Signal | Description |
|---|---|---|---|
| 1 | Brown | NC | No connect. Do not connect signals to this terminal. |
| 2 | White | AI- | Negative analog input signal and TEDS return. |
| 3 | Blue | NC | No connect. Do not connect signals to this terminal. |
| 4 | Black | AI+ | Positive analog input signal and TEDS data. |
| 5 | Gray | NC | No connect. Do not connect signals to this terminal. |
| * Wire color pertains to SHM125M I/O cables sold through NI. Other manufacturers' cable wire colors may vary. | |||
Power Connector Pinout
The following figure shows the pinout of the Power IN connector.
| Pin Number | Wire Color* | Signal | Description |
|---|---|---|---|
| 1 | Brown | V | Positive voltage line |
| 2 | White | Aux2 | Optional line for powering non-FieldDAQ devices |
| 3 | Blue | C | Common. Negative voltage line |
| 4 | Black | Aux1 | Optional line for powering non-FieldDAQ devices |
| 5 | Gray |
|
Chassis Ground. This terminal is internally connected to the C terminal. |
| * Wire color pertains to M125F power cables sold through NI. Other manufacturers' cable wire colors may vary. | |||
Ethernet Ports
The FD-11634 has two 8-pin X-coded M12 Ethernet ports—0 and 1.
You can use a shielded straight-through Ethernet or an Ethernet crossover cable with either of the Ethernet ports to network your device to a computer host, NI Linux Real-Time controller, another FieldDAQ device, or any network connection on the same subnet. Refer to Topology Options for more information about using these ports in various topologies.
| Pin Number | Wire Color | Gigabit Ethernet Signal | Fast Ethernet Signal |
|---|---|---|---|
| 1 | Orange/White | BI_DA+ | TX+ |
| 2 | Orange | BI_DA- | TX- |
| 3 | Green/White | BI_DB+ | RX+ |
| 4 | Green | BI_DB- | RX- |
| 5 | Brown/White | BI_DD+ | No Connect |
| 6 | Brown | BI_DD- | No Connect |
| 7 | Blue/White | BI_DC+ | No Connect |
| 8 | Blue | BI_DC- | No Connect |
You can use the Ethernet ports to reset the FieldDAQ device to factory-default settings. Refer to Resetting the FieldDAQ to Factory-Default Settings for more information.
Cap the Ethernet ports when not in use.
Cable Pinout
You can use the SHM125M-BNCF and SHM125M-BNCF-RA cables with the FD-11634.
The following figure shows the pinouts for the M12 and BNC connector on the cables.
| M12 Connector Pin Number | BNC Connector | Wire Color | Signal | Description |
|---|---|---|---|---|
| 1 | — | Brown | NC | No connect. Do not connect signals to this terminal. |
| 2 | BNC Shield | White | AI- | Negative analog input signal and TEDS return |
| 3 | — | Blue | NC | No connect. Do not connect signals to this terminal. |
| 4 | BNC Pin | Black | AI+ | Positive analog input signal and TEDS data |
| 5 | — | Gray | NC | No connect. Do not connect signals to this terminal. |
Refer to the SHM125M-BNCF Cable Safety, Environmental, and Regulatory Information or SHM125M-BNCF-RA Cable Safety, Environmental, and Regulatory Information for information about using these cables with your FieldDAQ device.
Input Characteristics
| Number of channels | 8 analog input channels |
| Isolation | Galvanic isolation between channels and to chassis |
| Input voltage range (AI+ to AI-) | ±10 V, ±1 V |
| ADC resolution | 24 bits |
| Type of ADC | Delta-Sigma (with analog prefiltering) |
| Sample mode | Simultaneous |
| Input coupling | Software-selectable AC/DC |
| TEDS support | IEEE 1451.4 TEDS Class I |
| TEDS capacitive drive | 5,000 pF |
| Frequency | 13.1072 MHz, 12.8 MHz, 12.288 MHz, 10.24 MHz |
| Accuracy | ±30 ppm maximum |
Base clocks can be synchronized with other FieldDAQ devices as well as CompactDAQ, CompactRIO, and other devices that support TSN network synchronization.
| Minimum | 500 Sample/s |
| Maximum | 102.4 kSample/s |
| Sampled data rates (fs) | Refer to the following table for sample data rates supported for each timebase |
| 13.1072 MHz | 12.8 MHz | 12.288 MHz | 10.24 MHz |
|---|---|---|---|
| 102.4 | 100.0 | 96.0 | 80.0 |
| 51.2 | 50.0 | 48.0 | 40.0 |
| 34.133 | 33.333 | 32.0 | 26.667 |
| 25.6 | 25.0 | 24.0 | 20.0* |
| 20.48 | 20.0 | 19.2 | 16.0 |
| 17.067 | 16.667 | 16.0* | 13.333 |
| 12.8 | 12.5 | 12.0 | 10.0* |
| 10.24 | 10.0 | 9.6 | 8.0 |
| 8.533 | 8.333 | 8.0* | 6.667 |
| 6.4 | 6.25 | 6.0 | 5.0* |
| 5.12 | 5.0 | 4.8 | 4.0 |
| 4.267 | 4.167 | 4.0* | 3.333 |
| 3.2 | 3.125 | 3.0 | 2.5* |
| 2.56 | 2.5 | 2.4 | 2.0 |
| 2.133 | 2.083 | 2.0* | 1.667 |
| 1.6 | 1.563 | 1.5 | 1.25* |
| 1.28 | 1.25 | 1.2 | 1.0 |
| 1.067 | 1.042 | 1.0* | 0.833 |
| 0.8 | 0.781 | 0.75 | 0.625 |
| 0.64 | 0.625 | 0.6 | 0.5 |
| Input impedance (AI+ to AI-) | 1 MΩ |
| Input capacitance (AI+ to AI-) | 520 pF |
| -3 dB | 0.53 Hz |
| -0.1 dB | 3.48 Hz |
| Nominal Input Range | Temperature | Gain Error (% of Reading) | DC-Coupled Offset Error (% of Range, mV) |
|---|---|---|---|
| ±10 V | 5 °C to 40 °C | 0.05%, typical | 0.012%, 1.2 mV, typical |
| 0.1%, maximum | 0.028%, 2.8 mV, maximum | ||
| -40 °C to 85 °C | 0.15%, maximum | 0.0078%, 7.8mV, maximum | |
| ±1 V | 5 °C to 40 °C | 0.06%, typical | 0.02%, 0.2 mV, typical |
| 0.12%, maximum | 0.04%, 0.4 mV, maximum | ||
| -40 °C to 85 °C | 0.2%, maximum | 0.14%, 1.4 mV, maximum |
| 5 °C to 40 °C | <5 mV typical |
| -40 °C to 85 °C | <50 mV typical |
| Input Range | Gain Drift | DC-Coupled Offset Drift |
|---|---|---|
| ±10 V | ±15 ppm/°C | ±50 μV/°C |
| ±1 V | ±20 ppm/°C | ±15 μV/°C |
| Gain mismatch (channel-to-channel, DC to 40 kHz) | 0.1 dB maximum |
| Phase mismatch (channel-to-channel, 1 kHz to 40 kHz) | 0.017°/kHz maximum |
| Phase nonlinearity (fs = 102.4 kSample/s, 1 kHz to 40 kHz) | 0.18° maximum |
| Crosstalk (1 kHz) | -120 dB |
| CMRR to chassis/earth (fin = 60 Hz) | 105 dB |
| Input Range | 1 kSample/s | 10 kSample/s | 102.4 kSample/s |
|---|---|---|---|
| ±10 V | 6.0 μV RMS | 9 μV RMS | 25 μV RMS |
| ±1 V | 0.7 μV RMS | 1.2 μV RMS | 3.5 μV RMS |
| Data Rate (kSample/s) | ADC Decimation Ratio | Input Range | |
|---|---|---|---|
| ±10 V | ±1 V | ||
| 102.4 | 64 | 108 | 106 |
| 51.2 | 128 | 111 | 109 |
| 25.6 | 256 | 114 | 112 |
| 12.8 | 512 | 117 | 115 |
| 6.4 | 1,024 | 120 | 118 |
| ±10 V input range | |
| ±1 V input range |
| Spurious Free Dynamic Range (SFDR), (1 kHz, -60 dBFS) | >130 dBFS |
| Input Range | 1 kHz | 20 Hz to 20 kHz | 20 kHz to 40 kHz |
|---|---|---|---|
| ±10 V | -105 dB | -98 dB | -90 dB |
| ±1 V | -105 dB | -88 dB | -75 dB |
| Input Range | 20 Hz to 20 kHz | 20 kHz to 40 kHz |
|---|---|---|
| ±10 V | -98 dB | -90 dB |
| ±1 V | -88 dB | -75 dB |
| Input Range | SMPTE 60 Hz + 7 kHz | CCIF 11 kHz + 12 kHz |
|---|---|---|
| ±10 V | -98 dB | -93 dB |
| ±1 V | -98 dB | -85 dB |
Test standards: SMPTE 60 Hz + 7 kHz, amplitude ratio 4:1 with total amplitude at 0 dBFS, and CCIF 11 kHz + 12 kHz, amplitude ratio 1:1 with each tone amplitude at -6 dBFS, up to 5th order harmonic.
IEPE
| Minimum | 4 mA |
| Typical | 4.17 mA |
| Excitation noise | 4 nA RMS, 0.1 Hz to 40 kHz BW |
| Detection threshold (AI+ to AI-) | 180 mV |
| Detection threshold hysteresis | 50 mV |
| Compliance voltage | 23 V maximum |
Filtering
Brickwall Filter (Default)
| Input delay | 36/fs + 1.5 μs |
| Input delay tolerance | ±100 ns |
| Passband frequency | DC to 0.4 · fs |
| 20 Hz to 20 kHz | ±0.03 dB maximum |
| 20 kHz to 40 kHz | ±0.05 dB maximum |
| Stopband frequency | At or above 0.5 · fs |
| Stopband rejection | ≥100 dB |
| Alias-free bandwidth | 0.5 · fs |
Butterworth Filter
| Input delay | Refer to the Butterworth Filter Input Delay for Available Timebases (fM) table. |
| Input delay tolerance | ±100 ns |
| Filter order | 2nd or 4th order |
| 13.1072 MHz | 12.8 MHz | 12.288 MHz | 10.24 MHz |
|---|---|---|---|
| 4,096 Hz | 4,000 Hz | 3,840 Hz | 3,200 Hz |
| 2,048 Hz | 2,000 Hz | 1,920 Hz | 1,600 Hz |
| 1,024 Hz | 1,000 Hz | 960 Hz | 800 Hz |
| 512 Hz | 500 Hz | 480 Hz | 400 Hz |
| 256 Hz | 250 Hz | 240 Hz | 200 Hz |
| 128 Hz | 125 Hz | 120 Hz | 100 Hz |
| Timebase (MHz) | Cutoff (Hz) | 4th Order | 2nd Order | ||
|---|---|---|---|---|---|
| Input Delay | Maximum Input Delay | Input Delay | Maximum Input Delay | ||
| 13.1072 | 4,096 | 436.0 μs | 457.4 μs | 398.7 μs | 405.1 μs |
| 2,048 | 537.1 μs | 580.3 μs | 453.3 μs | 466.3 μs | |
| 1,024 | 740.3 μs | 827.4 μs | 563.1 μs | 589.2 μs | |
| 512 | 1.1465 ms | 1.3208 ms | 782.9 μs | 834.8 μs | |
| 256 | 1.9583 ms | 2.3046 ms | 1.2224 ms | 1.3263 ms | |
| 128 | 3.5831 ms | 4.2770 ms | 2.0825 ms | 2.2924 ms | |
| 12.8 | 4.000 | 446.5 μs | 468.3 μs | 408.2 μs | 414.7 μs |
| 2.000 | 549.9 μs | 594.2 μs | 464.2 μs | 477.5 μs | |
| 1.000 | 758.1 μs | 847.2 μs | 576.5 μs | 603.3 μs | |
| 500 | 1.1739 ms | 1.3524 ms | 801.6 μs | 854.8 μs | |
| 250 | 2.0053 ms | 2.3598 ms | 1.2517 ms | 1.3581 ms | |
| 125 | 3.6691 ms | 4.3796 ms | 2.1324 ms | 2.3474 ms | |
| 12.288 | 3.840 | 465.0 μs | 487.8 μs | 425.1 μs | 432.0 μs |
| 1.920 | 572.8 μs | 618.9 μs | 483.5 μs | 497.3 μs | |
| 960 | 789.6 μs | 882.4 μs | 600.5 μs | 628.3 μs | |
| 480 | 1.2228 ms | 1.4087 ms | 834.9 μs | 890.4 μs | |
| 240 | 2.0888 ms | 2.4581 ms | 1.3038 ms | 1.4146 ms | |
| 120 | 3.8219 ms | 4.5620 ms | 2.2212 ms | 2.4451 ms | |
| 10.24 | 3.200 | 557.7 μs | 585.0 μs | 509.9 μs | 518.0 μs |
| 1,600 | 687.0 μs | 742.4 μs | 579.9 μs | 596.5 μs | |
| 800 | 947.2 μs | 1.0586 ms | 720.3 μs | 753.7 μs | |
| 400 | 1.4671 ms | 1.6902 ms | 1.0016 ms | 1.0682 ms | |
| 200 | 2.5063 ms | 2.9494 ms | 1.5643 ms | 1.6972 ms | |
| 100 | 4.5860 ms | 5.4741 ms | 2.6651 ms | 2.9338 ms | |
Comb Filter
| Input delay | 5/fs + 1.5 μs |
| Input delay tolerance | ±100 ns |
| Notches | fs, 2 fs, 3 fs, ... |
Time-Based Triggers
| Type | Start Trigger, Sync Pulse |
Timing and Synchronization
| Protocol | IEEE 802.1AS for network synchronization over 1000 Base-TX, full-duplex |
| Network synchronization accuracy[1]1 I/O synchronization is system-dependent. Assumes the devices are connected in a line topology. | <1 μs |
| Network synchronization accuracy with optimized configuration[2]2 I/O synchronization is system-dependent. Assumes a system containing one hop. | <100 ns |
For information about network synchronization accuracy, refer to NI-DAQmx-Based TSN Synchronization Accuracy Explained. For information about achieving high-accuracy synchronization, refer to How to Achieve High-Accuracy Measurements With NI-DAQmx-Based TSN Devices.
Network Interface
| Network protocols | TCP/IP, UDP |
| Network ports used | HTTP:80 (configuration only), TCP:3580; UDP:5353 (configuration only), TCP:5353 (configuration only); TCP:31415; UDP:7865 (configuration only), UDP:8473 (configuration only) |
| Network IP configuration | DHCP + Link-Local, DHCP, Static, Link-Local |
| Default MTU size | 1,500 bytes |
Ethernet
| Number of ports | 2 8-pin X-coded M12 ports, internally switched[3]3 This allows for line topologies or network redundancy. |
| Network interface | 1000 Base-TX, full-duplex; 1000 Base-TX, half-duplex; 100 Base-TX, full-duplex; 100 Base-TX, half-duplex; 10 Base-T, full-duplex; 10 Base-T, half-duplex |
| Communication rates | 10/100/1,000 Mbps, auto-negotiated |
| Maximum cabling distance | 100 m/segment |
| Maximum hops per line[4]4 With default software configuration. | 15 |
For information about creating reliable Ethernet-based systems, refer to Designing Distributed TSN Ethernet-Based Measurement Systems.
Power Requirements
| Vin | 9 V DC to 30 V DC |
| Vaux | Up to 30 V DC |
| Nominal | 7.3 W |
| Maximum | 10 W |
Current Limits
| Vin | 10 A maximum |
| Vaux | 10 A maximum total (combined with Vin) |
| Recommended external overcurrent protection | 16 A, slow blow fuse |
Physical Characteristics
| Dimensions | 198.5 mm × 77.4 mm × 47.1 mm (7.8 in. × 3.0 in. × 1.9 in.) |
| Weight | 1.179 kg (2 lb 9.6 oz) |
| Number | 8 |
| Type | 5-pin A-coded M12 connectors |
| Torque for M12 connectors (power, Ethernet, input connections) | 0.6 N · m (5.31 lb · in.) |
Calibration
| Calibration interval | 1 year |
Environmental Characteristics
Refer to the FD-11634 User Manual for more information about meeting these specifications.
| Operating | -40 °C to 85 °C |
| Storage | -40 °C to 100 °C |
| Operating | Up to 100% relative humidity, condensing or noncondensing |
| Storage | Up to 100% relative humidity, condensing or noncondensing |
| Ingress protection | IP65/IP67 |
| Pollution Degree | 4 |
| Maximum altitude | 5,000 m |
Shock and Vibration
| Random | 10 g RMS, 5 Hz to 2,000 Hz |
| Sinusoidal | 10 g, 5 Hz to 2,000 Hz |
| 12.4 mm minimum pk-pk displacement, 5 Hz to 20 Hz |
| Operating shock | 100 g, 11 ms half sine, 3 shocks at 6 orientations, 18 total |
| 40 g, 6 ms half sine, 4,000 shocks at 6 orientations, 24,000 total |
Environmental Standards
This product meets the requirements of the following environmental standards for electrical equipment.
- IEC 60068-2-1 Cold
- IEC 60068-2-2 Dry heat
- IEC 60068-2-6 Sinusoidal vibration
- IEC 60068-2-27 Shock
- IEC 60068-2-30 Damp heat, cyclic (12 h + 12 h cycle)
- IEC 60068-2-64 Broadband random vibration
Safety Voltages
The FD-11637 is rated for use in DRY or WET LOCATIONS. Do not connect hazardous voltages to the FD-11634. A hazardous voltage is a voltage greater than 30 V RMS, 42.4 V peak, or 60 V DC in DRY LOCATIONS and 22.6 V peak or 35 V DC in WET LOCATIONS.
Rated Voltages
Connect only voltages that are within the following limits:
| Between any two pins | 60 V DC (Dry Locations); 35 V DC (Wet Locations) |
Temporary Overvoltage Protection
Product has been designed to withstand power frequency overvoltage of relatively long duration as specified below. Voltages beyond these levels may cause permanent damage.
| Between any two pins on the connector | ±30 V DC |
Isolation Voltages
| Working Voltage | 60 V DC (Dry Locations); 35 V DC (Wet Locations) Non-Mains |
| Withstand | 1,000 V RMS, verified by 5 s withstand |
| Working Voltage | 60 V DC (Dry Locations); 35 V DC (Wet Locations) Non-Mains |
| Withstand | 1,000 V RMS, verified by 5 s withstand |
These test and measurement circuits are not rated for measurements performed on circuits directly connected to the electrical distribution system referred to as MAINS.
MAINS is a hazardous live electrical supply system to which equipment is designed to be connected to for the purpose of powering equipment. This product is rated for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.
Safety Compliance Standards
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1
Electromagnetic Compatibility Standards
This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
- EN 61326-1 (IEC 61326-1): Class A emissions; Industrial immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- AS/NZS CISPR 11: Group 1, Class A emissions
- ICES-001: Class A emissions
Product Certifications and Declarations
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
Related Information
- FD-11634 User Manual
- FD-11634 Calibration Procedure
- NI-DAQmx User Manual
- Software and Driver Downloads
- Release Notes
- License Setup and Activation
- Dimensional Drawings
- Product Certifications
- Letter of Volatility
- Discussion Forums
- NI Learning Center
- Topology Options
- Resetting the FieldDAQ to Factory-Default Settings
- Frequency Response of FieldDAQ Filters
- NI-DAQmx-Based TSN Synchronization Accuracy Explained
- How to Achieve High-Accuracy Measurements With NI-DAQmx-Based TSN Devices
- Designing Distributed TSN Ethernet-Based Measurement Systems