Set Current Limits
- Updated2025-11-03
- 1 minute(s) read
Setting Current Output Limits on a Fluke 5520A/5522A Calibrator
Complete the following steps to set current output limits on the Fluke 5520A and Fluke 5522A calibrators.
- Turn on the calibrator, and allow a warm-up period of at least 30 minutes.
- Press the RESET key.
- Press the SETUP key to open the Setup menu.
- To display the Instrument Setup menu, press INSTMT SETUP.
- To display the Output Setup menu, press OUTPUT SETUP.
- To display the Set Limits menu, press SET LIMITS.
- Press CURRENT.
- Press UPPER LIMIT, type the new Upper Limit value, and then press the ENTER key.
- Press LOWER LIMIT, type the new Lower Limit value, and then press the ENTER key.
- Press the PREV MENU key several times until the message STORE CHANGES/DISCARD CHANGES appears.
- Press STORE CHANGES.
- After the changes stored, press the RESET key.