Setting Current Output Limits on a Fluke 5520A/5522A Calibrator

Complete the following steps to set current output limits on the Fluke 5520A and Fluke 5522A calibrators.

  1. Turn on the calibrator, and allow a warm-up period of at least 30 minutes.
  2. Press the RESET key.
  3. Press the SETUP key to open the Setup menu.
  4. To display the Instrument Setup menu, press INSTMT SETUP.
  5. To display the Output Setup menu, press OUTPUT SETUP.
  6. To display the Set Limits menu, press SET LIMITS.
  7. Press CURRENT.
  8. Press UPPER LIMIT, type the new Upper Limit value, and then press the ENTER key.
  9. Press LOWER LIMIT, type the new Lower Limit value, and then press the ENTER key.
  10. Press the PREV MENU key several times until the message STORE CHANGES/DISCARD CHANGES appears.
  11. Press STORE CHANGES.
  12. After the changes stored, press the RESET key.