Calibrating NI-9227 Using Calibration Executive
- Updated2025-11-03
- 3 minute(s) read
Calibrate the NI-9227 using Calibration Executive.
Calibration Executive Procedure Features
| Verify Only |
Verify & Adjust |
Manual Mode |
Optional Adjust |
Approximate Test Time
| Verify Only | Verify & Adjust |
|---|---|
| 3 minutes | 8 minutes |
Test Equipment
The following table lists the test equipment required for calibrating NI-9227 modules.
| Instrument | Recommended Model | Where Used | Requirements |
|---|---|---|---|
| Calibrator | Fluke 5520A | Accuracy, Adjustment |
Use a high-precision current source with an accuracy ≤100 ppm. |
| Function Generator | NI 5421 | Gain Matching, Phase Matching |
Use a function generator with an output impedance of 50Ω capable of sourcing both 1 kHz and 20 kHz ±10 Vpk AC sinusoidal signals with amplitude uncertainty of ±10% to a 50Ω load. |
| Termination Resistor | — | Gain Matching, Phase Matching |
50 Ω, 1 W, ±5% |
| Chassis | cDAQ-9178 | All | — |
Note To avoid an error, the NI-9227 calibration procedure requires
that you set current limits of the calibrator to
more than ±5 Amps.
Test Conditions
The following setup and environmental conditions are required to ensure the NI-9227 meets published specifications.
- Keep connections to the device as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements.
- Verify that all connections to the device are secure.
- Use shielded copper wire for all cable connections to the device. Use twisted-pairs wire to eliminate noise and thermal offsets.
- Maintain an ambient temperature of 23 °C ± 5 °C. The device temperature will be greater than the ambient temperature.
- Keep relative humidity below 80%.
- Allow a warm-up time of at least 10 minutes to ensure that the NI-9227 measurement circuitry is at a stable operating temperature.
Device Setup
- Make sure the cDAQ-9178 power source is not connected.
- Connect the cDAQ-9178 to the system safety ground.
- Attach a ring lug to a 14 AWG (1.6 mm) wire.
- Connect the ring lug to the ground terminal on the side of the cDAQ-9178 using the ground screw.
- Attach the other end of the wire to the system safety ground.
- Install the module in slot 8 of the cDAQ-9178 chassis. Leave slots 1 through 7 of the cDAQ-9178 chassis empty.
- Connect the cDAQ-9178 chassis to your host computer.
- Connect the power source to the cDAQ-9178 chassis.
- Launch Measurement & Automation Explorer (MAX).
- Right-click the device name and select Self-Test to ensure that the module is working properly.
Test Limit Equations
The following test limits are derived from the published specifications.
| Equation Type | Equation |
|---|---|
| Accuracy | TestLimits = TestValue ± (TestValue *
GainError (% of Reading) + OffsetError)
|
| Gain Matching | ±0.13 dB Max |
| Phase Matching | ±0.1 Deg/kHz Max |