NI-9157/9159 Explosive Atmospheres User Guide
- Updated2026-01-14
- 3 minute(s) read
NI-9157/9159 Explosive Atmospheres User Guide
Product Descriptions
NI-9157: 14-Slot, Virtex-5 LX85 FPGA, MXI-Express CompactRIO Chassis
NI-9159: 14-Slot, Virtex-5 LX110 FPGA, MXI-Express CompactRIO Chassis
Physical Characteristics
| |||||||
Dimensions | Visit ni.com/dimensions and search by model number. | ||||||
| |||||||||||||
| |||||||||||||
Environmental
Operating temperature | 0 °C to 55 °C |
Storage temperature | -40 °C to 85 °C |
Ingress protection | IP40 |
Operating humidity | 10% RH to 90% RH, noncondensing |
Storage humidity | 5% RH to 95% RH, noncondensing |
Pollution degree | 2 |
Maximum altitude | 2,000 m |
Indoor use only.
Hazardous Locations
Explosive atmospheres rating | Ex ec IIC T4 Gc |
CCC certificate number | 2021312310000310 |
Safety Guidelines
Operate the product only as described in this document.
Safety Guidelines for Hazardous Locations
The NI-9157/9159 have been evaluated as Ex ec IIC T4 Gc equipment and are CCC certified. Each product is suitable for use within ambient temperatures of 0 °C ≤ Ta ≤ 55 °C in either nonhazardous locations or Zone 2 hazardous locations.
Follow these guidelines if you are installing the product in a potentially explosive environment. Not following these guidelines may result in serious injury or death.
Safety Voltages
Connect only voltages that are below these limits.
V terminal to C terminal | 30 V DC maximum, Measurement Category I |
Power Requirements
NI-9157
Voltage requirement | 9 V to 30 V | ||||||
Recommended power supply | 55 W, 30 V DC maximum | ||||||
| |||||||
NI-9159
Voltage requirement | 90 V to 30 V | ||||||
Recommended power supply | 55 W, 30 V DC maximum | ||||||
| |||||||
Installing C Series Modules
Verify that power is not connected to the I/O connector(s) on the C Series module.
Removing C Series Modules
Verify that power is not connected to the I/O connector(s) on the C Series module before you remove a module from the chassis.
Product Certifications and Declarations
To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
Worldwide Support and Services
NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504, USA.