Each issue appears as a row in the table and includes the following fields:
ID | Known Issue | |||||
---|---|---|---|---|---|---|
689967 | WTS Software may fail to start up if the WTS Instrument Software License for Qualcomm Connectivity Interface Support is activated without an NI-RFmx WLAN license. The WTS Instrument Software License for Qualcomm Connectivity Interface Support also requires a valid NI-RFmx WLAN license. You must purchase and activate this license separately. Failure to activate the license can cause WTS Software to report errors at startup and not operate as expected. Workaround: Ensure you purchase and activate the NI-RFmx WLAN license.
| |||||
675405 | In WTS Software, the instrument is unresponsive after sending a misformatted string using raw sockets. The string must be sent using one of the following ways:
Workaround: Use the same pling at the beginning and end of the strings. For example:
| |||||
677407 | In WTS Software, the PXIe-5840 has a high LO leakage that causes measurement degradation when using Bluetooth. When using Bluetooth with the PXIe-5840, the LO leakage is strong enough to degrade the measurements. The numbers may be fine for production test, but for characterization they may not be suitable. Workaround: Contact your NI representative to request a fix.
| |||||
679212 | In WTS Software, auto broadcasting might fail during the first few signal generations. When executing scenarios with auto broadcasting enabled, clients might not be joined or might throw an error during the first one to two signal generations, depending on how the signal is generated. This is related to the first time execution speed of some commands after the clients are initially joined by calling the SYSTem:RFSG:CONCurrency:TESTrun Start command, and also depends on the configured join time. Workaround: Perform either of the following options as a workaround.
| |||||
677464 | In WTS Software, the WTS Instrument Interface terminates when the software restarts or is shutdown. The WTS Instrument Interface requires a data connection to the WTS Software. When the connection to the WTS Software is lost, either because the instrument restarts or the software is terminated, all connected local and remote instances of the WTS Instrument Interface are terminated. Workaround: Disconnect all instances of the WTS Instrument Interface before issuing a restart command.
| |||||
651101 | WTS Software does not work with other regional settings other than English-US. SCPI and VISA support only using a period (.) as a decimal point and comma (,) to separate parameters and lists. WTS Software malfunctions and becomes unusable if you use any other settings. Workaround: Ensure you are using (.) as the decimal symbol value, and (,) as the list separator value in Regional Settings»Additional Settings.
| |||||
583918 | In WTS Software, the WTS scheduling chart displays the signal analyzer as active when it waits for a trigger. The signal analyzer is active when waiting for a trigger, and you cannot use it for anything else during that time. Workaround: N/A
|
Related Topics
TestStand Wireless Test Module and Wireless Test System 1.4 and 1.4.1 Known Issues
Contact NI regarding this document or issues in the document. If you contact NI in regards to a specific issue, reference the ID number given in the document. The ID number contains the current issue ID number as well as the legacy ID number (use the current ID number when contacting NI). You can contact us through any of the normal support channels including phone, email, or the discussion forums. Visit the NI Website to contact us. Also contact us if you find a workaround for an issue that is not listed in the document.