在物理通道名称中,<产品名称>-<序列号>是默认的设备名称。例如,序列号为1C6A39C的FD-11613在MAX中显示为FD11613-1C6A39C。MAX还会添加组编号。例如,FD11613-1C6A39C-Bank1。在MAX中可以更改该名称。

FieldDAQ设备包括FD-11601、FD-11603、FD-11605、FD-11613、FD-11614、FD-11634和FD-11637。

模拟输入

下表列出用于FieldDAQ设备的模拟输入物理通道的数量和名称。

设备 通道数 名称
FD-11601 8 FD-11601-xxxxxxx-Bank1/ai0FD-11601-xxxxxxx-Bank1/ai3

FD-11601-xxxxxxx-Bank2/ai0FD-11601-xxxxxxx-Bank2/ai3

FD-11603 8 FD-11603-xxxxxxx-Bank1/ai0FD-11603-xxxxxxx-Bank1/ai3

FD-11603-xxxxxxx-Bank2/ai0FD-11603-xxxxxxx-Bank2/ai3

FD-11605 8 FD-11605-xxxxxxx-Bank1/ai0FD-11605-xxxxxxx-Bank1/ai3

FD-11605-xxxxxxx-Bank2/ai0FD-11605-xxxxxxx-Bank2/ai3

FD-11613 8 FD-11613-xxxxxxx-Bank1/ai0 FD-11613-xxxxxxx-Bank1/ai7
FD-11614 16 FD-11614-xxxxxxx-Bank1/ai0FD-11614-xxxxxxx-Bank1/ai7

FD-11614-xxxxxxx-Bank2/ai0FD-11614-xxxxxxx-Bank2/ai7

FD-11634 8 FD-11634-xxxxxxx-Bank1/ai0FD-11634-xxxxxxx-Bank1/ai3

FD-11634-xxxxxxx-Bank2/ai0FD-11634-xxxxxxx-Bank2/ai3

FD-11637 8 FD-11637-xxxxxxx-Bank1/ai0FD-11637-xxxxxxx-Bank1/ai3

FD-11637-xxxxxxx-Bank2/ai0FD-11637-xxxxxxx-Bank2/ai3

应变和惠斯通桥测量FD-11637

FD-11637设备仅支持AI应变计、AI花型应变计、AI力桥、AI压强桥、AI扭矩桥、AI桥(V/V)和带激励的AI自定义电压等通道类型。

将FD-11637设备与带激励的AI自定义电压通道一起使用时,必须将AI.Excit.UseForScaling属性设置为True。该属性将导致通道返回比值数据:Vin/Vex。

FD-11637设备将返回电压比率而不是电压。因此,通过AI.Bridge.InitialRatio属性指定初始电压比率,或者将AI.Bridge.InitialVoltage属性设置为设备返回的比率Vin/Vex,乘以激励电压Vex。

当FD-11637处于1/4桥模式时,需使用AI.Bridge.NomResistance属性控制通道使用120 Ω范围或350 Ω范围。

FD-11637具有内部半桥,以及用于120 Ω和350 Ω应变计的1/4桥。