LCR:AC Stimulus:Current Amplitude
- Updated2025-10-06
- 1 minute(s) read
Specifies the amplitude, in A RMS, of the AC current test signal applied to the DUT for LCR measurements.
This property applies when the LCR Stimulus Function property is set to AC Current .
Note
This property is not supported by all instruments. Refer to
Supported Properties by Device
for information about supported instruments.
Valid Values : 7.08e-9 A RMS to 0.707 A RMS
Instrument specifications affect the valid values you can program. Refer to the specifications for your instrument for more information.
Default Value : Refer to Supported Properties by Device for the default value by instrument.
Remarks
The following table lists the characteristics of this property.
| Short Name | LCR Current Amplitude |
| Data type | ![]() |
| Permissions | Read/Write |
| High-level VIs | N/A |
| Channel-based | Yes |
| Instrument-based | No |
| Resettable | No |