Defines a user waveform for either Standard Function or Frequency List output mode.

To select the waveform, set the Waveform parameter of the niFgen Configure Standard Waveform VI or niFgen Create Frequency List VI to User.

The waveform data must be scaled between -1.0 and 1.0. Use the Amplitude parameter of the niFgen Configure Standard Waveform VI to generate different output voltages.

Note You must set the Output Mode to Standard Function or Frequency List using the niFgen Configure Output Mode VI before calling this VI.


icon

Inputs/Outputs

  • cstr.png Channel Name

    Channel Name specifies the channel that this VI uses.

  • civrn.png Instrument Handle

    Instrument Handle identifies a particular instrument session.

    Instrument Handle is obtained from the niFgen Initialize VI, niFgen Initialize With Options VI, or the niFgen Initialize With Channels VI.

  • ci32.png Waveform Size

    Waveform Size specifies the size of the waveform in samples.

  • c1ddbl.png Waveform Data Array

    Waveform Data Array specifies the array of data you want to load into the new waveform.

    You must normalize the data points in the array to be between -1.00 and +1.00.

  • cerrcodeclst.png error in (no error)

    error in describes error conditions that occur before this node runs. This input provides standard error in functionality.

    Default value: no error

  • iivrn.png Instrument Handle Out

    Instrument Handle Out passes a reference to your instrument session to the next VI.

  • ierrcodeclst.png error out

    error out contains error information. This output provides standard error out functionality.