PXI-2530B Specifications

PXI-2530B Specifications

Caution The protection provided by the PXI-2530B can be impaired if it is used in a manner not described in this document.
Notice To ensure the specified EMC performance, operate this product only with shielded cables and accessories.

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.

Specifications are Warranted unless otherwise noted.

Conditions

Specifications are valid at 23 °C unless otherwise noted.

All voltages are specified in DC, ACpk, or a combination unless otherwise specified.

Topology

Topologies[1]1 The PXI-2530B has eight interconnected banks of 1-wire 16 × 1 multiplexers. These multiplexers can be used in any combination with the independent topology.

1-wire 128 x 1 multiplexer

1-wire dual 64 x 1 multiplexer

1-wire quad 32 x 1 multiplexer

1-wire octal 16 x 1 multiplexer

2-wire 64 x 1 multiplexer

2-wire dual 32 x 1 multiplexer

2-wire quad 16 x 1 multiplexer

4-wire 32 x 1 multiplexer

4-wire dual 16 x 1 multiplexer

1-wire 4 x 32 matrix

1-wire 8 x 16 matrix

2-wire 4 x 16 matrix

Independent

Input Characteristics

Caution Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are for other circuits not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV. Do not connect the PXI-2530B to signals or use for measurements within Measurement Categories II, III, or IV.
Maximum switching voltage[2]2 Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.

Channel-to-channel

60 VDC, 30 VACrms, CAT I

Channel-to-ground

60 VDC, 30 VACrms, CAT I

Maximum current (per terminal or internal path)

0.4 A

Maximum switching power (per channel, resistive)

10 W

DC path resistance[3]3 Path resistance is a combination of relay contact resistance and trace resistance. Contact resistance typically remains low for the life of a relay. At the end of relay life, the contact resistance may rapidly rise above 3 Ω.(channel-to-common)

Initial

<2 Ω, typical

End-of-life

≥3 Ω, typical

Thermal EMF (1-wire configuration, channel-to-common)

<50 µV, typical

RF Performance Characteristics

Bandwidth (50 Ω system, 1-wire configuration referenced to 1WREF)

16 × 1

>19 MHz, typical

128 × 1

>12 MHz, typical

Channel-to-channel isolation (50 Ω system, 1-wire configuration referenced to 1WREF)

200 kHz

>60 dB, typical

1 MHz

>40 dB, typical

Dynamic Characteristics

Simultaneous drive limit

64 relays

Relay operate/release time[4]4 Certain applications may require additional time for proper settling.

555 µs

Relay life (no load)[5]5 Reed relays are highly susceptible to damage caused by switching capacitive and inductive loads. Capacitive loads can cause high inrush currents while inductive loads can cause high flyback voltages. The addition of appropriate resistive protection can greatly improve contact lifetime. For more information about adding protection circuitry to a capacitive load, visit ni.com/info and enter the Info Code relaylifetime. For information about inductive loads, enter the Info Code relayflyback. [6]6 The relays used in the PXI-2530B are field replaceable.

Mechanical

1 × 109 cycles, typical

Electrical (resistive, <10 pF load)

10 V, 100 mA

1 × 108 cycles, typical

25 V, 400 mA

5 × 106 cycles, typical

60 V, 160 mA

1 × 106 cycles, typical

Trigger Characteristics

Input trigger

Sources

PXI trigger lines <0…7>, front panel

Minimum pulse width[7]7 The PXI-2530B can recognize trigger pulse widths less than 150 ns if digital filtering is disabled.

150 ns

Front panel input voltage

-0.5 V, minimum

+0.7 V, VIL maximum

+2.0 V, VIH minimum

+3.3 V, nominal

+5.5 V, maximum

Output trigger

Destinations

PXI trigger lines <0…7>, front panel

Default pulse width

2 µs, typical

Front panel voltage

+3.3 V TTL, 8 mA, nominal

Physical Characteristics

Relay type[8]8 NI advises against installing reed relay modules directly adjacent to an embedded controller with a magnetic hard drive because of the sensitivity of reed relays and the possibility of interference.

Reed

Front panel connector

LFH matrix 50, 160 positions, female

PXI power requirement

5 V

4.2 W

3.3 V

1.2 W

Dimensions (L × W × H)

3U, one slot, PXI/cPCI module

21.6 × 2.0 × 13.0 cm (8.5 × 0.8 × 5.1 in.)

Weight

266 g (9.4 oz)

Environment

Operating temperature

0 °C to 55 °C

Storage temperature

-40 °C to 70 °C

Relative humidity

5% to 85%, noncondensing

Pollution Degree

2

Maximum altitude

2,000 m

Indoor use only.

Shock and Vibration

Table 1. Shock and Vibration
Operational Shock 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.)
Random vibration, operating 5 Hz to 500 Hz, 0.3 grms
Random vibration, nonoperating 5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-2-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.)

Compliance and Certifications

Safety Compliance Standards

This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:

  • IEC 61010-1, EN 61010-1
  • UL 61010-1, CSA C22.2 No. 61010-1
Note For safety certifications, refer to the product label or the Product Certifications and Declarations section.

CE Compliance

This product meets the essential requirements of applicable European Directives, as follows:

  • 2014/35/EU; Low-Voltage Directive (safety)
  • 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
  • 2011/65/EU; Restriction of Hazardous Substances (RoHS)
  • 2014/53/EU; Radio Equipment Directive (RED)
  • 2014/34/EU; Potentially Explosive Atmospheres (ATEX)

Product Certifications and Declarations

Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.

Environmental Management

NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

EU and UK Customers

  • Waste Electrical and Electronic Equipment (WEEE)—At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.
  • 电子信息产品污染控制管理办法(中国RoHS)

  • 中国RoHSNI符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于NI中国RoHS合规性信息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)
  • 1 The PXI-2530B has eight interconnected banks of 1-wire 16 × 1 multiplexers. These multiplexers can be used in any combination with the independent topology.

    2 Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.

    3 Path resistance is a combination of relay contact resistance and trace resistance. Contact resistance typically remains low for the life of a relay. At the end of relay life, the contact resistance may rapidly rise above 3 Ω.

    4 Certain applications may require additional time for proper settling.

    5 Reed relays are highly susceptible to damage caused by switching capacitive and inductive loads. Capacitive loads can cause high inrush currents while inductive loads can cause high flyback voltages. The addition of appropriate resistive protection can greatly improve contact lifetime. For more information about adding protection circuitry to a capacitive load, visit ni.com/info and enter the Info Code relaylifetime. For information about inductive loads, enter the Info Code relayflyback.

    6 The relays used in the PXI-2530B are field replaceable.

    7 The PXI-2530B can recognize trigger pulse widths less than 150 ns if digital filtering is disabled.

    8 NI advises against installing reed relay modules directly adjacent to an embedded controller with a magnetic hard drive because of the sensitivity of reed relays and the possibility of interference.