BTS-16110 Specifications

BTS-16110 Specifications

These specifications apply to the BTS-16110.

Revision History

Version Date changed Description
378999B-01 November 2025 Correcting typo in unit of measurement for input current ranges.
378999A-01 May 2024 Initial release.

Looking For Something Else?

For information not found in the specifications for your product, such as operating instructions, browse Related Information.

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
  • Measured specifications describe the measured performance of a representative model.

Specifications are Typical unless otherwise noted.

Conditions

Specifications are valid under the following conditions unless otherwise noted.

  • Operating temperature of 15 °C to 40 °C guaranteed, 40 °C to 50 °C extended operating range
  • Recommended calibration interval of 2 years. The BTS-16110 will not meet specifications unless operated within the recommended calibration interval
  • Background calibration enabled
  • 30 minutes warm-up time before operation

Physical

Table 1. Physical specifications
Dimensions (L×W×H) 327 mm × 147 mm × 83 mm (12.94 in. × 5.80 in. × 3.27 in.)
Weight 3.63 kg (8.00 lb)

Input Characteristics

Table 2. General Input Characteristics
Measurement signals Voltage and current
Internal data rate 1.25 MS/s
Internal ADC resolution 18 bits
Input voltage range ±1,500 V DC
Input current ranges ±10 A DC, ±2 A DC, ±667 mA DC
Data rates 1.25 MS/s, 1 kS/s
Data format 64-bit, double precision
-3 dB bandwidth, 1.25 MHz data rate 60 kHz, typical
-3 dB bandwidth, 1 kHz data rate 1 kHz, typical
Table 3. Data Rates
1.25 MS/s Voltage and current
1 kS/s 1 ms average voltage and current
Table 4. Nominal Voltages
HI to LO ±1,500 V DC, maximum
HI to Earth ±1,000 V DC, maximum
LO to Earth ±1,000 V DC, maximum

Accuracy Specifications

Table 5. DC Voltage ± (% of Reading + mV)
Range Input Resistance Two Year at 15 °C to 40 °C
1,500 V DC 10 MΩ

0.03 + 100, maximum

0.006 + 10, typical

Table 6. DC Current ± (% of Reading + % of Range)
Range Input Resistance Two Year at 15 °C to 40 °C
10 V DC >1 GΩ

0.01 + 0.001, maximum

0.0025 + 0.0003, typical

2 A DC 0.5 Ω

0.05+ 0.005, maximum

0.015 + 0.0015, typical

667 mA DC 0.5 Ω

0.05 + 0.0075, maximum

0.015 + 0.0015, typical

Note Ambient temperature variations faster than 1 °C/min may disrupt the BTS-16110 thermal regulation.
Note Forced airflow over the BTS-16110 reduces its ability to thermoregulate at low temperatures and low input voltages.
Note Warranted specifications are valid from 15 °C to 40 °C. The device can operate at the non warranted specifications from 40 °C to 50 °C.
Note Prolonged operation at temperatures above 40 °C might cause device accuracy to shift outside of warranted specifications and require recalibration to restore accuracy specifications. Take all precautions to maintain ambient air temperatures within guaranteed accuracy specifications.

Input Noise

Table 7. DC Voltage Noise
1 kS/s data rate with no averaging 1 mV RMS, typical
1.25 MS/s data rate

75 mV RMS with background calibration enabled, typical

15 mV RMS with background calibration disabled, typical

Table 8. DC Current, Voltage Input Noise
1 kS/s data rate 5 µV RMS, typical
1.25 MS/s data rate 120 µV RMS, typical
Table 9. DC Current, Amps Input Noise
1 kS/s data rate 1.2 µA (2 A)/0.8 µA (667 mA) RMS, typical
1.25 MS/s data rate 30 µA (2 A)/15 µA (667 mA) RMS, typical

Network/Ethernet Ports

Table 10. General network specifications
Number of ports 2
Network interface 10Base-T, 100Base-TX, 1000Base-T Ethernet
Compatibility IEEE 802.3
Communication rates 10 Mb/s, 100 Mb/s, 1000 Mb/s auto-negotiated
Maximum cabling distance 100 m/segment

Network Timing and Synchronization

Table 11. Network timing and synchronization
Protocol IEEE 802.1AS
Network synchronization accuracy <1 μs

USB Ports

Table 12. USB type-C device data port
USB interface USB 2.0, Hi-Speed
Maximum data rate 480 Mb/s
Maximum current (from host) 250 mA
Table 13. USB type-C host display port
USB interface USB 3.1 Gen1, SuperSpeed
Maximum data rate 5 Gb/s
Maximum current 900 mA
Alternate mode DisplayPort

Power Requirements

Table 14. Power requirements
Voltage input range 19 V DC to 30 V DC
Input power consumption Up to 150 W, maximum

Safety Voltages

Connect only voltages that are below these limits.

Table 15. Safety voltages
Input voltage range ±1,500 V DC
Input current ranges ±10 V DC, ±2 A DC, ±667 mA DC

Measurement Category

Table 16. Measurement category
CAT O and CAT I ±1,000 V DC
Caution Do not connect the product to signals or use for measurements within Measurement Categories II, III, or IV.
Note Measurement Categories CAT O and CAT I are equivalent. These test and measurement circuits are for other circuits not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.

Environmental Characteristics

Table 17. Temperature specifications
Warranted specification operating 15 °C to 40 °C
Extended operating 40 °C to 50 °C
Storage -40 °C to 85 °C
Table 18. Humidity specifications
Operating 15% RH to 75% RH, noncondensing
Storage 5% RH to 95% RH, noncondensing
Table 19. General environmental specifications
Pollution Degree 2
Maximum altitude 2,000 m
Table 20. Shock and vibration specifications
Operating vibration 5 Hz to 500 Hz, 0.3 g RMS
Non-operating vibration 5 Hz to 500 Hz, 2.4 g RMS
Operating shock 30 g, half-sine, 11 ms pulse

Calibration

Table 21. Calibration
Recommended warm-up time 30 minutes
Calibration interval 2 years