BTS-16110 Specifications
- Updated2025-11-24
- 4 minute(s) read
BTS-16110 Specifications
BTS-16110 Specifications
These specifications apply to the BTS-16110.
Revision History
| Version | Date changed | Description |
|---|---|---|
| 378999B-01 | November 2025 | Correcting typo in unit of measurement for input current ranges. |
| 378999A-01 | May 2024 | Initial release. |
Looking For Something Else?
For information not found in the specifications for your product, such as operating instructions, browse Related Information.
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
- Measured specifications describe the measured performance of a representative model.
Specifications are Typical unless otherwise noted.
Conditions
Specifications are valid under the following conditions unless otherwise noted.
- Operating temperature of 15 °C to 40 °C guaranteed, 40 °C to 50 °C extended operating range
- Recommended calibration interval of 2 years. The BTS-16110 will not meet specifications unless operated within the recommended calibration interval
- Background calibration enabled
- 30 minutes warm-up time before operation
Physical
| Dimensions (L×W×H) | 327 mm × 147 mm × 83 mm (12.94 in. × 5.80 in. × 3.27 in.) |
| Weight | 3.63 kg (8.00 lb) |
Input Characteristics
| Measurement signals | Voltage and current |
| Internal data rate | 1.25 MS/s |
| Internal ADC resolution | 18 bits |
| Input voltage range | ±1,500 V DC |
| Input current ranges | ±10 A DC, ±2 A DC, ±667 mA DC |
| Data rates | 1.25 MS/s, 1 kS/s |
| Data format | 64-bit, double precision |
| -3 dB bandwidth, 1.25 MHz data rate | 60 kHz, typical |
| -3 dB bandwidth, 1 kHz data rate | 1 kHz, typical |
| 1.25 MS/s | Voltage and current |
| 1 kS/s | 1 ms average voltage and current |
| HI to LO | ±1,500 V DC, maximum |
| HI to Earth | ±1,000 V DC, maximum |
| LO to Earth | ±1,000 V DC, maximum |
Accuracy Specifications
| Range | Input Resistance | Two Year at 15 °C to 40 °C |
|---|---|---|
| 1,500 V DC | 10 MΩ |
0.03 + 100, maximum 0.006 + 10, typical |
| Range | Input Resistance | Two Year at 15 °C to 40 °C |
|---|---|---|
| 10 V DC | >1 GΩ |
0.01 + 0.001, maximum 0.0025 + 0.0003, typical |
| 2 A DC | 0.5 Ω |
0.05+ 0.005, maximum 0.015 + 0.0015, typical |
| 667 mA DC | 0.5 Ω |
0.05 + 0.0075, maximum 0.015 + 0.0015, typical |
Input Noise
| 1 kS/s data rate with no averaging | 1 mV RMS, typical |
| 1.25 MS/s data rate |
75 mV RMS with background calibration enabled, typical 15 mV RMS with background calibration disabled, typical |
| 1 kS/s data rate | 5 µV RMS, typical |
| 1.25 MS/s data rate | 120 µV RMS, typical |
| 1 kS/s data rate | 1.2 µA (2 A)/0.8 µA (667 mA) RMS, typical |
| 1.25 MS/s data rate | 30 µA (2 A)/15 µA (667 mA) RMS, typical |
Network/Ethernet Ports
| Number of ports | 2 |
| Network interface | 10Base-T, 100Base-TX, 1000Base-T Ethernet |
| Compatibility | IEEE 802.3 |
| Communication rates | 10 Mb/s, 100 Mb/s, 1000 Mb/s auto-negotiated |
| Maximum cabling distance | 100 m/segment |
Network Timing and Synchronization
| Protocol | IEEE 802.1AS |
| Network synchronization accuracy | <1 μs |
USB Ports
| USB interface | USB 2.0, Hi-Speed |
| Maximum data rate | 480 Mb/s |
| Maximum current (from host) | 250 mA |
| USB interface | USB 3.1 Gen1, SuperSpeed |
| Maximum data rate | 5 Gb/s |
| Maximum current | 900 mA |
| Alternate mode | DisplayPort |
Power Requirements
| Voltage input range | 19 V DC to 30 V DC |
| Input power consumption | Up to 150 W, maximum |
Safety Voltages
Connect only voltages that are below these limits.
| Input voltage range | ±1,500 V DC |
| Input current ranges | ±10 V DC, ±2 A DC, ±667 mA DC |
Measurement Category
| CAT O and CAT I | ±1,000 V DC |
Environmental Characteristics
| Warranted specification operating | 15 °C to 40 °C |
| Extended operating | 40 °C to 50 °C |
| Storage | -40 °C to 85 °C |
| Operating | 15% RH to 75% RH, noncondensing |
| Storage | 5% RH to 95% RH, noncondensing |
| Pollution Degree | 2 |
| Maximum altitude | 2,000 m |
| Operating vibration | 5 Hz to 500 Hz, 0.3 g RMS |
| Non-operating vibration | 5 Hz to 500 Hz, 2.4 g RMS |
| Operating shock | 30 g, half-sine, 11 ms pulse |
Calibration
| Recommended warm-up time | 30 minutes |
| Calibration interval | 2 years |