Sweep:Segment:Test Receiver Attenuation Enabled
- Updated2025-10-13
- 1 minute(s) read
Specifies whether VNA performs measurements using the same test receiver attenuation for all segments or uses different test receiver attenuation for each segment. This property is used only when you set the Sweep Type property to Segment.
You do not need to use a selector string to configure or read this property for the default signal instance. Refer to the Selector Strings topic for information about the string syntax for named signals.
The default value is False.
Remarks
The following table lists the characteristics of this property.
| Short Name | Segment Test Receiver Attn Enabled |
| Data type | ![]() |
| Permissions | Read/Write |
| High-level VIs | N/A |
| Resettable | Yes |
| False | 0 | All segments are measured with the test receiver attenuation that you specify using Test Receiver Attenuation property. |
| True | 1 | The selected segment is measured with the test receiver attenuation that you specify using Segment Test Receiver Attenuation property. |
