Correction:Calibration:Calkit:Type
- Updated2025-10-13
- 1 minute(s) read
Specifies the type of calkit used for calibration.
Use "port::all" as the selector string to specify the same calkit type for all VNA ports. Use "port::<portname>" as the selector string to read this property for a specific VNA port.
Note Note that, currently, RFmxVNA only supports the use of same calkit type (and the same calkit) for all calibration ports in a guided calibration procedure.
The default value is Electronic.
Remarks
The following table lists the characteristics of this property.
| Short Name | Calkit Type |
| Data type | ![]() |
| Permissions | Read/Write |
| High-level VIs | N/A |
| Resettable | Yes |
| Electronic | 0 | Perform calibration using an electronic calkit module. Supported Electronic Calkit Modules: NI CAL-5501. |
| Mechanical | 1 | Perform calibration using discrete cal standards from a mechanical calkit. |