Correction:Port Subset:Enabled
- Updated2025-10-13
- 1 minute(s) read
Specifies whether to enable correction for a subset of ports for which calibration data is avaialble in the calset. This property is used only when you set the Correction Enabled property to True.
Enable this property to exclude subset of calset ports from measurement error correction, to achieve faster measurement speed by avoiding extra acquisitions. Use Correction Port Subset Full Ports and Correction Port Subset Response Ports to specify the list of subset ports to be included in the measurement error correction.
The default value is False.
Remarks
The following table lists the characteristics of this property.
| Short Name | Correction Port Subset Enabled |
| Data type | ![]() |
| Permissions | Read/Write |
| High-level VIs | N/A |
| Resettable | Yes |
| False | 0 | The measurement disables port-subsetting for error correction. |
| True | 1 | The measurement enabes port-subsetting for error correction. |