Initializing Instruments and Opening a New Session in the RTG Software
- Updated2025-05-05
- 4 minute(s) read
Initializing Instruments and Opening a New Session in the RTG Software
Initialize the instruments you want to use and open a new session. Initializing moves RTG from the Idle state to the Inactive state.
- Run RTG UI.exe from the <Program Files>\National
Instruments\RTG\app folder on the host computer. The RTG User
Interface opens.Note The server must be running for RTG to receive gRPC commands.
- In the instrument tree, select the instrument(s) to use for the session. You can initialize
more than one VST instrument in a session. Only supported instruments that have
been configured in NI MAX appear in the instrument tree.
Figure 43. RTG User Interface
Personality Menu Options Menu Utilities Menu Help Menu Instrument Tree RTG State Initialize Session Button Start/Stop Button Close Session Button - Click the Personality menu and select Pulse Invariant or Pulse Aware.
- Click Initialize Session to initialize all selected instruments.
- If using mmWave radio heads with the PXIe-5831/5832, click the Options menu, select the instrument, and click Enable mmWave Heads.
- If you want to disable frequency correction, click the Options menu, select the instrument, and click Enable Frequency Correction. Frequency correction is enabled by default.
- Click Configuration under each initialized instrument to open the
instrument configuration pane.
Figure 44. Configuration Pane
Center Frequency Field Pulse Repetition Interval Field Reference Level Field Hardware Trigger Source Selector Recommended LO In Frequency Indicator Recommended LO In Power Indicator Enable/Disable External LO Button LO In Frequency Selector LO In Power Selector - Enter the value for the VST in the Center Frequency field for each initialized instrument.
- Enter the value for the VST in the Reference Level field for each initialized instrument. Refer to Reference Level and Pulse Detection Calibration for more information about configuring this field.
- If using predictive mode, enter the Pulse Repetition Interval.
- If using a hardware trigger, select a hardware trigger port in the Hardware
Trigger Source selector:
- PFI0
- PXI_Trig0
- PXI_Trig1
- PXI_Trig2
- PXI_Trig3
- PXI_Trig4
- PXI_Trig5
- PXI_Trig6
- PXI_Trig7
Note The hardware trigger source, also known as external trigger in the API, is configured as pxi_trigger_input in the INI file. - Click Calibration under each initialized instrument to
access the Calibration pane.
Figure 45. RTG Calibration Pane Minimum Time Delay Indicator External Time Delay Field Offset Frequency Field Minimum Attenuation Indicator Output External Attenuation Field Target Common Attenuation Field - Update the following settings to manually account for any offset, attenuation,
and delay not included in the system self-calibration:
- External Time Delay
- Output External Attenuation
- Offset Frequency
- Target Common Attenuation
Related Information
- Reference Level and Pulse Detection Calibration
It is recommended that the radar signal levels be adjusted to a ±10 dBm range.