Factory Default Settings (Initialization)
- Updated2025-04-02
- 2 minute(s) read
Factory Default Settings (Initialization)
If you hold down SHIFT+LOCAL while you turn the POWER switch on, the RMX programmable power supply will be reset to the factory default settings.
The factory default settings are shown in the following table.
|
Basic Item |
Parameter Description |
|---|---|
|
Output Voltage |
0 V |
|
Output Current |
105% of the rated output current |
|
OVP (Overvoltage Protection) |
112% of the rated output current |
|
OCP (Overcurrent Protection) |
112% of the rated output current |
|
UVL (Undervoltage Limit) |
0 V |
|
Parameter Number |
CONFIG Parameter |
Parameter Description |
|---|---|---|
|
CF01 |
Master-slave parallel operation parameter |
1 (independent operation) |
|
CF021 |
Power-on status parameter |
SAFE (panel settings that were in use the last time the power was turned off) |
|
CF03 |
Method for clearing OHP, FAN, and AC-FAIL alarms |
SAFE (do not clear the alarm automatically) |
|
CF041 |
Memory content display parameter |
ON (displayed) |
|
CF051 |
CC control using an Vext or Rext |
OFF |
|
CF061 |
CV control using an Vext or Rext |
OFF |
|
CF071 |
CV/CC control range parameter |
LO (0 to 5 V) |
|
CF08 |
Range during voltage and current monitoring |
LO (0 to 5 V) |
|
CF091 |
External control parameter for turning output on and off |
OFF |
|
CF101 |
External control logic parameter for turning output on and off |
LO (a LOW signal turns output on)2 |
|
CF11 |
Bleeder on/off parameter |
ON (bleeder circuit is on) |
|
CF121 |
Output-on startup state parameter |
CV (CV is prioritized) |
|
CF13 |
Detection time of OCP activation |
0 ms |
|
CF141 |
Current setting limit |
ON (the setting is limited) |
|
CF151 |
Voltage setting limit |
ON (the setting is limited) |
|
CF16 |
Panel display brightness setting |
7 |
|
CF17 |
Communication monitor timer setting |
OFF (Communication monitoring timer off) |
|
CF18 |
Unused |
|
|
CF193 |
Display setting on slave units for master-slave parallel operation |
OFF (not displayed on slave units) |
|
CF20 |
LAN interface parameter |
LAN (the LAN interface will be used) |
|
CF21 |
Command language parameter |
SCPI |
|
CF22 |
Emulation parameter during remote control |
NONE (RMX programmable power supply commands) |
|
CF23 |
SCPI communication error display parameter |
OFF (not displayed) |
|
CF24 |
Multichannel (VMCB) domain number |
0 (the multichannel function is not used) |
|
CF25 |
Multichannel (VMCB) channel number |
0 (the RMX programmable power supply is used as a master unit on the multichannel network) |
|
CF41 |
IP address assignment method |
DHCP: ONAUTO IP: ONMANUAL IP: OFF |
|
CF47 |
SCPI-RAW protocol TCP port number |
5025 |
|
CF52 |
RS232C baud rate parameter |
19.2 (kbit/s) |
|
1 You can only specify this parameter when the RMX programmable power supply is being used as a master unit.2 0 to 0.5 V or shorted.3 You can only specify this parameter when the RMX programmable power supply is being used as a slave unit. |