Factory Default Settings (Initialization)

If you hold down SHIFT+LOCAL while you turn the POWER switch on, the RMX programmable power supply will be reset to the factory default settings.

The factory default settings are shown in the following table.

Table 50. Table 2-9. Basic Settings

Basic Item

Parameter Description

Output Voltage

0 V

Output Current

105% of the rated output current

OVP (Overvoltage Protection)

112% of the rated output current

OCP (Overcurrent Protection)

112% of the rated output current

UVL (Undervoltage Limit)

0 V

Table 51. Table 2-10. CONFIG Settings

Parameter Number

CONFIG Parameter

Parameter Description

CF01

Master-slave parallel operation parameter

1 (independent operation)

CF021

Power-on status parameter

SAFE (panel settings that were in use the last time the power was turned off)

CF03

Method for clearing OHP, FAN, and AC-FAIL alarms

SAFE (do not clear the alarm automatically)

CF041

Memory content display parameter

ON (displayed)

CF051

CC control using an Vext or Rext

OFF

CF061

CV control using an Vext or Rext

OFF

CF071

CV/CC control range parameter

LO (0 to 5 V)

CF08

Range during voltage and current monitoring

LO (0 to 5 V)

CF091

External control parameter for turning output on and off

OFF

CF101

External control logic parameter for turning output on and off

LO (a LOW signal turns output on)2

CF11

Bleeder on/off parameter

ON (bleeder circuit is on)

CF121

Output-on startup state parameter

CV (CV is prioritized)

CF13

Detection time of OCP activation

0 ms

CF141

Current setting limit

ON (the setting is limited)

CF151

Voltage setting limit

ON (the setting is limited)

CF16

Panel display brightness setting

7

CF17

Communication monitor timer setting

OFF (Communication monitoring timer off)

CF18

Unused

CF193

Display setting on slave units for master-slave parallel operation

OFF (not displayed on slave units)

CF20

LAN interface parameter

LAN (the LAN interface will be used)

CF21

Command language parameter

SCPI

CF22

Emulation parameter during remote control

NONE (RMX programmable power supply commands)

CF23

SCPI communication error display parameter

OFF (not displayed)

CF24

Multichannel (VMCB) domain number

0 (the multichannel function is not used)

CF25

Multichannel (VMCB) channel number

0 (the RMX programmable power supply is used as a master unit on the multichannel network)

CF41

IP address assignment method

DHCP: ONAUTO IP: ONMANUAL IP: OFF

CF47

SCPI-RAW protocol TCP port number

5025

CF52

RS232C baud rate parameter

19.2 (kbit/s)