What to Use

  • RM-26999
  • DAQ Multifunction I/O Module appropriate for your system configuration
  • Device-appropriate cable
  • 24 V power supply
  • Power connector, included in kit

What to Do

Complete the following steps to connect the RM-26999 to a DAQ Multifunction I/O Module.

Note Before you connect your RM-26999, install, connect, and configure all PXI chassis, PXI controllers, and modules appropriate for your desired system configuration. Refer to the DAQ for PXI/PXI Express Getting Started for the PXI Express module installation and connection procedures.
Figure 16. Connecting the RM-26999 to a DAQ Multifunction I/O Module


  1. Verify that the protective conductor terminals are connected to protective earth ground.
  2. Connect the device-appropriate cable to the RM-26999 and secure it using the retention screws.
  3. Connect the device-appropriate cable to the DAQ Multifunction I/O Module.
  4. Connect a 24 V power supply to the power terminal on the RM-26999.
  5. Power on the DAQ Multifunction I/O Module.
  6. Power on the 24 V power supply.
  7. Verify that the RM-26999 Status LED is on.