Specifies whether the device is the master device when synchronizing the Sample Clock between multiple devices. The master device distributes the Sample Clock sync signal to all devices in the system through the Sample Clock sync distribution line.

When synchronizing the Sample Clock, one device must always be designated as the master. The master device actively drives the Sample Clock sync distribution line.

To set this property, the NI-RFSG device must be in the Configuration state.

Supported Devices: PXIe-5646

Default Value: FALSE

TRUE The device is the master device for synchronizing the Sample Clock.
FALSE The device is not the master device for synchronizing the Sample Clock.

Related Topics

Synchronization Using NI-RFSA and NI-RFSG—Refer to this topic for more information about PXIe-5646 device synchronization.

Remarks

The following table lists the characteristics of this property.

Short Name Sync Sample Clock Master
Data type cbool.png
Permissions Read/Write
High-level VIs N/A
Channel-based No
Resettable Yes