Arb:Waveform Attributes:Waveform RF Blanking
- Updated2025-10-14
- 1 minute(s) read
Enables or disables RF blanking.
| RF Blanking Source | RF Blanking | Behaviour |
|---|---|---|
| "" (empty string) | Disable | No blanking performed. |
| "" (empty string) | Enable | Blanking performed based on burst start and stop values and blanking source set to private marker. |
| marker0, marker1, marker2, or marker3 | Disable | Blanking performed based on the marker locations for the marker that the user set in the blanking source. |
| marker0, marker1, marker2, or marker3 | Enable | Error is shown. |
Supported Devices: PXIe-5820/5830/5831/5832/5840/5841/5842
Note For PXIe-5830/5831/5832, the RF Blanking reserves a PXI trigger line. If you are calling any
niRFSG Reset or
niRFSA Reset on the same device, NI recommends calling it before committing blanking properties. Alternatively, you can call
niRFSG Reset With Options or
niRFSA Reset With Options. Select Routes in the steps to omit parameter.
Default Values: Disabled
Remarks
The following table lists the characteristics of this property.
| Short Name | Waveform RF Blanking |
| Data type | ![]() |
| Permissions | Read/Write |
| High-level VIs | N/A |
| Channel-based | Yes |
| Resettable | Yes |
| Disable | 0 | RF blanking is disabled. |
| Enable | 1 | RF blanking is enabled. |