RF:ALC Control
- Updated2025-10-14
- 1 minute(s) read
Enables or disables the automatic leveling control (ALC).
PXIe-5654 with PXIe-5696: If this property is enabled, the ALC is closed (closed-loop mode) and allows for better amplitude accuracy and wider amplitude dynamic range. If this property is disabled, the ALC is open (open-loop mode), which is ideal when using modulation. Disabling the ALC Control property also allows NI-RFSG to perform an automatic power search.
PXIe-5654: Disable is the only supported value for this device. The PXIe-5654 does not support the ALC when used as a stand-alone device.
Supported Devices: PXIe-5654/5654 with PXIe-5696
Default Values:
PXIe-5654: Disable
PXIe-5654 with PXIe-5696: Enable
Related Topics
ALC Closed Loop Versus Open Loop
Remarks
The following table lists the characteristics of this property.
| Short Name | ALC Control |
| Data type | ![]() |
| Permissions | Read/Write |
| High-level VIs | N/A |
| Channel-based | No |
| Resettable | Yes |
| Disable | 0 | Disables the ALC. |
| Enable | 1 | Enables the ALC. |