RF:Amplitude Settling
- Updated2025-10-14
- 1 minute(s) read
Configures the amplitude settling accuracy in decibels. NI-RFSG waits until the RF power settles within the specified accuracy level after calling the niRFSG Initiate VI or niRFSG Wait Until Settled VI or prior to advancing to next step if using RF list mode.
Any specified amplitude settling value that is above the acceptable minimum value is coerced down to the closest valid value.
PXI/PXIe-5650/5651/5652: This property is for NI internal use only.
Supported Devices: PXIe-5654/5654 with PXIe-5696, PXIe-5820/5830/5831/5832/5840/5841/5842/5860
Default Values:
PXIe-5654: 4
PXIe-5654 with PXIe-5696 (ALC disabled): 4
PXIe-5654 with PXIe-5696 (ALC enabled): 0.2
PXIe-5820/5830/5831/5832/5840/5841/5842/5860: 0.5
Valid Values:
PXIe-5654: 1.5, 2, 4
PXIe-5654 with PXIe-5696 (ALC disabled): 1.5, 2, 4
PXIe-5654 with PXIe-5696 (ALC enabled): 0.2, 0.5
PXIe-5820/5830/5831/5832/5840/5841/5842/5860: 0.01 to 1
Related Topics
Remarks
The following table lists the characteristics of this property.
| Short Name | Amplitude Settling |
| Data type | ![]() |
| Permissions | Read/Write |
| High-level VIs | N/A |
| Channel-based | No |
| Resettable | Yes |