Sweep:Pulse:Acquisition:Auto
- Updated2025-10-13
- 1 minute(s) read
Specifies whether, based on Pulse Modulator Width, the measurement automatically sets appropriate Pulse Acquisition Delay and Pulse Acquisition Width. This property is used only when you set Pulse Mode Enabled property to True.
You do not need to use a selector string to configure or read this property for the default signal instance. Refer to the Selector Strings topic for information about the string syntax for named signals.
The default value is True.
Remarks
The following table lists the characteristics of this property.
| Short Name | Pulse Acquisition Auto |
| Data type | ![]() |
| Permissions | Read/Write |
| High-level VIs | N/A |
| Resettable | Yes |
| False | 0 | The Pulse Acquisition Delay and Pulse Acquisition Width properties are not automatically set by the measurement. The values that you set for these properties are used by the measurement. |
| True | 1 | The measurement uses the Pulse Modulator Width property to automatically set appropriate values for Pulse Acquisition Delay and Pulse Acquisition Width properties. Pulse Acquisition Delay is set as the sum of Pulse Modulator Delay and approximately 20% of the Pulse Modulator Width, and Pulse Acquisition Width is set to approximately 75% of the Pulse Modulator Width. |