Groups

Group members

NameDescription
RFMXSPECAN_ATTR_AMPM_ALL_TRACES_ENABLED

Specifies whether to enable the traces to be stored and retrieved after performing the AMPM measurement.

RFMXSPECAN_ATTR_AMPM_AM_TO_AM_ENABLED

Specifies whether to enable the results that rely on the AM to AM characteristics.

RFMXSPECAN_ATTR_AMPM_AM_TO_PM_ENABLED

Specifies whether to enable the results that rely on AM to PM characteristics.

RFMXSPECAN_ATTR_AMPM_DUT_AVERAGE_INPUT_POWER

Specifies the average power of the signal at the input port of the device under test. This value is expressed in dBm.

RFMXSPECAN_ATTR_AMPM_EVM_ENABLED

Specifies whether to enable the RFMXSPECAN_ATTR_AMPM_RESULTS_MEAN_RMS_EVM attribute.

RFMXSPECAN_ATTR_AMPM_FREQUENCY_OFFSET_CORRECTION_ENABLED

Enables frequency offset correction for the measurement.

RFMXSPECAN_ATTR_AMPM_IQ_ORIGIN_OFFSET_CORRECTION_ENABLED

Enables IQ origin offset correction for the measurement.

RFMXSPECAN_ATTR_AMPM_MAXIMUM_TIMING_ERROR

Specifies the maximum time alignment error expected between the acquired and the reference waveforms. This value is expressed in seconds.

RFMXSPECAN_ATTR_AMPM_MEASUREMENT_ENABLED

Specifies whether to enable the AMPM measurement.

RFMXSPECAN_ATTR_AMPM_MEASUREMENT_INTERVAL

Specifies the duration of the reference waveform considered for the AMPM measurement. When the reference waveform contains an idle duration, the AMPM measurement neglects the idle samples in the reference waveform leading up to the start of the first active portion of the reference waveform. This value is expressed in seconds.

RFMXSPECAN_ATTR_AMPM_NUMBER_OF_ANALYSIS_THREADS

Specifies the maximum number of threads used for parallelism for the AMPM measurement.

RFMXSPECAN_ATTR_AMPM_REFERENCE_POWER_TYPE

Specifies the reference power used for AM to AM and AM to PM traces.

RFMXSPECAN_ATTR_AMPM_SIGNAL_TYPE

Specifies whether the reference waveform is a modulated signal or a combination of one or more sinusoidal signals. To time-align the sinusoidal reference waveform to the acquired signal, set the AMPM Signal Type attribute to Tones, which switches the AMPM measurement alignment algorithm.

RFMXSPECAN_ATTR_AMPM_SYNCHRONIZATION_METHOD

Specifies the method used for synchronization of acquired waveform with reference waveform.