Advanced:Self Calibration:Self Cal Validity Check
- Updated2025-10-13
- 1 minute(s) read
Specifies whether the RFmx driver validates the self-calibration data.
You can specify the time interval required to perform the check using the Self Cal Validity Check Time Interval property.
NI recommends to perform self-calibration using the RFmxInstr Self Calibrate VI when RFmx reports an invalid self-calibration data warning.
The default value is Off.
Supported devices: PXIe-5663/5663E/5665/5668, PXIe-5644/5645/5646, PXIe-5820/5830/5831/5832/5833/5840/5841/5842/5860
Remarks
The following table lists the characteristics of this property.
| Short Name | Self Cal Validity Check |
| Data type | ![]() |
| Permissions | Read/Write |
| High-level VIs | N/A |
| Resettable | Yes |
| Off | 0 | Indicates that RFmx does not check whether device self-calibration data is valid. |
| Enabled | 1 | Indicates that RFmx checks whether device self-calibration data is valid and reports a warning from the RFmx Commit and RFmx Initiate VIs when the data is invalid. |