NI-TClk Manual Calibration Using NI-RFSG
- Updated2023-02-20
- 1 minute(s) read
NI-TClk Manual Calibration Using NI-RFSG
Complete the following steps to improve the synchronization accuracy by lowering skew and minimizing jitter of NI-TClk synchronized devices using manual calibration.
- Connect the RF output connector on the vector signal transceivers (VSTs) to a multi-channel signal analyzer. For best performance, minimize unnecessary skew by using matched length cables.
- Configure the VST for generation using NI-TClk. If using LabVIEW, use the example.
- Start the generation and measure the skew between devices, noting the amount of jitter.
- Read the value from the Absolute Delay property or NIRFSG_ATTR_ABSOLUTE_DELAY attribute. Stop the generation, modify the value, and write it back to apply skew correction.
- Repeat steps 3 and 4 until you achieve the appropriate synchronization results.
- Read the value from the Absolute Delay property or NIRFSG_ATTR_ABSOLUTE_DELAY attribute, and store the value for future synchronization sessions.