All of the input and output connections on the PXI-4110 front panel have user-replaceable fuses.
Refer to the table below for fuse ratings and manufacturer information.
| Input/Output | Fuse Rating | Description | Recommended Manufacturer Part Number |
|---|---|---|---|
| Channels 0, 1, 2 | F 1.5 A 125 V | User-replaceable chip fuse | Littelfuse 045301.5 |
| Auxiliary power input | T 6.3 A L 250 V | User-replaceable 5 × 20 mm glass fuse | Littelfuse 21806.3 |
Tip Although blown
output channel and auxiliary power input fuses are not software detectable, you can
use the niDCPower Auxiliary Power Source Available property or the
NIDCPOWER_ATTR_AUXILIARY_POWER_SOURCE_AVAILABLE attribute to troubleshoot the
auxiliary power input fuse.
Identify the fuse you want to replace using the following figure.
- Output Channel Fuse (Channel 0)
- Output Channel Fuse (Channel 2)
- Output Channel Fuse (Channel 1)
- Auxiliary Input Fuse
- Spare Output Channel Fuse
To replace a fuse on the PXI-4110, complete the following steps:
- Power down the chassis.
- Disconnect all output and auxiliary power connections.
- Remove the PXI-4110 from the chassis.
-
Remove the fuse.
Fuse Process Output channel fuse - Using small pliers, gently pull the fuse to release it from the fuse holder.
Auxiliary input power fuse - Using a flathead screwdriver, turn the fuse holder cap counter-clockwise to release it from the PXI-4110 front panel.
- Gently pull the auxiliary power input fuse to release it from the fuse holder cap.
-
Install the replacement fuse.
Fuse Process Output channel fuse - Using small pliers, gently place the replacement fuse into the fuse holder.
Auxiliary input power fuse Use the auxiliary input fuse holder to service the auxiliary fuse on the PXI-4110. - Slide the replacement fuse into the fuse holder cap.
- Screw the cap clockwise to replace it on the PXI-4110 front panel.
Note You can use the spare fuse included on the PXI-4110 to replace an output channel fuse.