PXI-2720 Specifications
- Updated2023-02-18
- 5 minute(s) read
PXI-2720 Specifications
This document lists specifications for the PXI-2720 (PXI-2720) 8-bit resistor module. All specifications are subject to change without notice. Visit ni.com/manuals for the most current specifications.
About These Specifications
Specifications characterize the warranted performance of the instrument under the stated operating conditions.
Typical Specifications are specifications met by the majority of the instrument under the stated operating conditions and are tested at 23 °C ambient temperature. Typical specifications are not warranted. The following specifications are typical at 23 °C unless otherwise specified.
All voltages are specified in DC, ACpk, or a combination unless otherwise specified.
Topology | Independent |
Refer to the NI Switches Help at ni.com/manuals for detailed topology information.
Input Characteristics
Channel accuracy is a function of the overall resistor accuracy, plus relay and trace path resistance, plus any effects of temperature.
Typical minimum resistance | 1 Ω | ||||||
|
Channel resistance is a combination of relay contact resistance, trace resistance, and resistor value. Contact resistance typically remains low for the life of a relay. At the end of relay life, the offset resistance may rapidly rise above 3 Ω.
The following figures show the expected resistance error and the expected resistance.
Dynamic Characteristics
Simultaneous drive limit | 89 relays |
Relay operate/release time | 250 μs |
|
Trigger Characteristics
|
Physical Characteristics
Relay type | Reed |
Front panel connector | 37-pin D-SUB connector |
DMM port | 2 × 2 Micro-Fit connector |
Power requirement | 1.2 W at 3.3 V 11 W maximum at 5 V |
Dimensions (L × W × H) | 3U, one slot, PXI/cPCI module, 21.6 × 2.0 × 13.0 cm (8.5 × 0.8 × 5.1 in.) |
Weight | 266 g (9.3 oz) |
Environment
Operating temperature | 0 °C to 55 °C |
Storage temperature | -20 °C to 70 °C |
Relative humidity | 5% to 85%, noncondensing |
Pollution Degree | 2 |
Maximum altitude | 2,000 m |
Indoor use only.
Shock and Vibration
Operational Shock | 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) | ||||||
|
Compliance and Certifications
Safety
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA 61010-1
Electromagnetic Compatibility
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- AS/NZS CISPR 11: Group 1, Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
CE Compliance

This product meets the essential requirements of applicable European Directives, as follows:
- 2014/35/EU; Low-Voltage Directive (safety)
- 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
Online Product Certification
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
Waste Electrical and Electronic Equipment (WEEE)
电子信息产品污染控制管理办法(中国RoHS)
1 Measurement Categories CAT I and CAT O (Other) are equivalent. These test and measurement circuits are not intended for direct connection to the MAINs building installations of Measurement Categories CAT II, CAT III, or CAT IV.