Configuring a S11 1 Coupler System
- Updated2025-05-13
- 1 minute(s) read
Complete the following steps to configure your hardware for S-parameter measurements using S11 one coupler per DUT port.
Note The PXIe-5831 does not have dedicated RF
IN/OUT ports. Use the IF IN/OUT 1 port as RF IN and the IF IN/OUT 0 port as RF OUT when
using a PXIe-5831.
- Connect the RF OUT or IF IN/OUT 0 port of the out port of the coupler using SMA cable.
- Connect the coupled port to the RF IN or IF IN/OUT 1 port.
- Connect the input port of the coupler to the DUT for measurement or the calibration kit for calibration using SMA cables.
- If using the PXIe-5841, share the local oscillator (LO) signal by connecting the TX LO OUT port to the RX LO IN port of the VST using a MMPX-to-MMPX cable. If using the PXIe-5831, no additional connection needs to be made as the LO is shared internally.
Related Information
- S11 1 Coupler System Hardware Requirements
In this hardware configuration, use one coupler on the port of the DUT to measure the reflection wave of the DUT.