RfsaAdvancedVertical Class
- Updated2025-10-14
- 2 minute(s) read
Represents the advanced properties related to vertical.
Derives from
Syntax
Namespace: NationalInstruments.ModularInstruments.NIRfsa
public class RfsaAdvancedVertical : RfsaSubObject
Remarks
For more information, refer to NI RF Vector Signal Analyzers Help.
Properties
| Name | Description |
|---|---|
| AmplitudeSettling | Gets or sets the amplitude settling accuracy in decibels. |
| DeviceConfigurationTemperature | Gets or sets the temperature, in degrees Celsius, that NI-RFSA uses to calculate the device configuration settings. |
| DigitalGain | Gets or sets the scaling factor, in decibels, applied to the time-domain voltage data in the digitizer. NI-RFSA does not compensate for the specified digital gain. NI-RFSA does not compensate for the specified digital gain. |
| ExternalGain | Gets or sets the gain of a switch or cable, in dB, connected before the RF IN connector of an NI-RFSA device. |
| MechanicalAttenuation | Gets or sets the level of mechanical attenuation, in dB, for the RF path. |
| MinimumAcpr | Gets or sets the minimum adjacent channel power ratio (ACPR), in dB, relative to the main channel reference level. |
| NI5663 | Gets the object used to access the advanced vertical properties specific to NI 5663. |
| NI5665 | Gets the object used to access the advanced vertical properties specific to NI 5665. |
| NI5693 | Gets the object used to access the advanced vertical properties specific to NI 5693. |
| NI5694 | Gets the object used to access the advanced vertical properties specific to NI 5694. |
| OspDataScalingFactor | Gets or sets the scaling factor applied to the time-domain voltage data in the IF digitizer. |
| OverflowErrorReporting | Gets or sets error reporting for ADC and onboard signal processing overflows. Overflows lead to clipping of the waveform. |
| ReferenceLevelHeadroom | Gets or sets the margin NI-RFSA adds to the ReferenceLevel property. The margin helps to avoid clipping and overflow warnings if the input signal exceeds the configured reference level. |
| RFAttenuation | Gets or sets the downconverter module attenuation setting. |
| RFAttenuationStepSize | Gets or sets the step size for the RF attenuation level. |
| RFPreampEnabled | Gets or sets whether to enable RF preamplifier on the downconverter. |
| ThermalCorrectionHeadroomRange | Gets or sets the expected thermal operating range of the instrument from the self-calibration temperature, in degrees Celsius, returned from GetDeviceTemperature. |
| ThermalCorrectionTemperatureResolution | Gets or sets the temperature change, in degrees Celsius, required before NI-RFSA recalculates the thermal correction settings when entering the Running state. |