Represents the advanced properties related to vertical.

Derives from

Syntax

Namespace: NationalInstruments.ModularInstruments.NIRfsa

public class RfsaAdvancedVertical : RfsaSubObject

Remarks

For more information, refer to NI RF Vector Signal Analyzers Help.

Properties

NameDescription
AmplitudeSettling

Gets or sets the amplitude settling accuracy in decibels.

DeviceConfigurationTemperature

Gets or sets the temperature, in degrees Celsius, that NI-RFSA uses to calculate the device configuration settings.

DigitalGain

Gets or sets the scaling factor, in decibels, applied to the time-domain voltage data in the digitizer. NI-RFSA does not compensate for the specified digital gain. NI-RFSA does not compensate for the specified digital gain.

ExternalGain

Gets or sets the gain of a switch or cable, in dB, connected before the RF IN connector of an NI-RFSA device.

MechanicalAttenuation

Gets or sets the level of mechanical attenuation, in dB, for the RF path.

MinimumAcpr

Gets or sets the minimum adjacent channel power ratio (ACPR), in dB, relative to the main channel reference level.

NI5663

Gets the object used to access the advanced vertical properties specific to NI 5663.

NI5665

Gets the object used to access the advanced vertical properties specific to NI 5665.

NI5693

Gets the object used to access the advanced vertical properties specific to NI 5693.

NI5694

Gets the object used to access the advanced vertical properties specific to NI 5694.

OspDataScalingFactor

Gets or sets the scaling factor applied to the time-domain voltage data in the IF digitizer.

OverflowErrorReporting

Gets or sets error reporting for ADC and onboard signal processing overflows. Overflows lead to clipping of the waveform.

ReferenceLevelHeadroom

Gets or sets the margin NI-RFSA adds to the ReferenceLevel property. The margin helps to avoid clipping and overflow warnings if the input signal exceeds the configured reference level.

RFAttenuation

Gets or sets the downconverter module attenuation setting.

RFAttenuationStepSize

Gets or sets the step size for the RF attenuation level.

RFPreampEnabled

Gets or sets whether to enable RF preamplifier on the downconverter.

ThermalCorrectionHeadroomRange

Gets or sets the expected thermal operating range of the instrument from the self-calibration temperature, in degrees Celsius, returned from GetDeviceTemperature.

ThermalCorrectionTemperatureResolution

Gets or sets the temperature change, in degrees Celsius, required before NI-RFSA recalculates the thermal correction settings when entering the Running state.