Closes an NI-RFSA external calibration session and, if specified, stores the new calibration constants and calibration data, such as time and temperature, in the onboard EEPROM.

Supported Devices: PXIe-5601/5603/5605/5606 (external digitizer mode), PXIe-5693/5694/5698, PXIe-5820/5830/5831/5832/5840/5841/5842/5860


icon

Inputs/Outputs

  • civrn.png instrument handle

    instrument handle identifies your calibration instrument session. instrument handle is obtained from the niRFSA Init Ext Cal VI.

  • cu16.png action

    action specifies how to use the calibration values from this session as the session is closed.

    Abort (1500)

    The old calibration constants are kept, and the new ones are discarded. This is the default behavior.

    Commit (1501)

    The new calibration constants are stored in the EEPROM.

  • cerrcodeclst.png error in

    error in describes error conditions that occur before this node runs. This input provides standard error in functionality.

  • ierrcodeclst.png error out

    error out contains error information. So if an error was contained in error in and this VI also produces an error, error out contains the same information as error in. For all other cases, this output provides standard error out functionality.