Triggering:Trigger Level Property
- Updated2023-02-21
- 1 minute(s) read
Short Name: Trigger Level
Property of niScope
Specifies the voltage threshold for the trigger. The units are volts.
This property affects instrument behavior only when the Trigger Type is set to Edge, Hysteresis, or Window.
The values of the range and offset parameters in niScope Configure Vertical determine the valid range for the trigger level on the channel you use as the trigger source. The value you pass for this parameter must meet the following conditions:
Trigger Level <= Vertical Range/2 + Vertical Offset
Trigger Level >= (-Vertical Range/2) + Vertical Offset
Remarks
The following table lists the characteristics of this property.
Datatype | ![]() |
Permissions | Read/Write |
High-level VIs | niScope Configure Trigger Edge (Analog Edge Ref Trigger), niScope Configure Trigger Glitch (Glitch Trigger), niScope Configure Trigger Hysteresis (Analog Hysteresis Ref Trigger), niScope Configure Trigger Width (Width Trigger) |
Channel-based | No |
Resettable | No |