Triggering:Trigger Hysteresis Property
- Updated2023-02-21
- 1 minute(s) read
Short Name: Trigger Hysteresis
Property of niScope
Specifies the size of the hysteresis window, in volts, on either side of the trigger level.
The digitizer triggers when the trigger signal passes through the threshold you specify with the Trigger Level parameter, has the slope you specify with the Trigger Slope parameter, and passes through the hysteresis window that you specify with this property.
Default Value: 0.05
Valid Values
| Positive trigger slope | 0 <= Hysteresis <= Trigger Level + (Vertical Range / 2) - Vertical Offset |
| Negative trigger slope | 0 <= Hysteresis <= (Vertical Range / 2) + Vertical Offset - Trigger Level |
Remarks
The following table lists the characteristics of this property.
| Datatype | ![]() |
| Permissions | Read/Write |
| High-level VIs | niScope Configure Trigger Hysteresis (Analog Hysteresis Ref Trigger) |
| Channel-based | No |
| Resettable | No |