NI-SCOPE Properties

Triggering:Trigger Runt:Runt Time Condition Property

  • Updated2023-02-21
  • 1 minute(s) read

Short Name: Runt Time Condition

Property of niScope

Specifies whether runt triggers are time qualified, and if so, how the oscilloscope triggers in relation to the duration range bounded by the Runt Time Low Limit and Runt Time High Limit properties.

Default Value: None

None (0)

Time qualification is disabled. The oscilloscope triggers on runt pulses based solely on the voltage level of the pulses.

Within (1)

The oscilloscope triggers on pulses that, in addition to meeting runt voltage criteria, have a duration within the range bounded by Runt Time Low Limit and Runt Time High Limit.

Outside (2)

The oscilloscope triggers on pulses that, in addition to meeting runt voltage criteria, have a duration not within the range bounded by Runt Time Low Limit and Runt Time High Limit.

Remarks

The following table lists the characteristics of this property.

Datatype iL
PermissionsRead/Write
High-level VIs N/A
Channel-based No
ResettableNo