Short Name: End of Record to Advance Trigger Holdoff

Property of niScope

End of Record to Advance Trigger Holdoff is the length of time (in seconds) that a device waits between the completion of one record and the acquisition of pre-trigger samples for the next record. During this time, the acquisition engine state delays the transition to the Wait for Advance Trigger state, and will not store samples in onboard memory, accept an Advance Trigger, or trigger on the input signal..

Supported Devices: NI 5185/5186

Remarks

The following table lists the characteristics of this property.

Datatype fD
PermissionsRead/Write
High-level VIs N/A
Channel-based No
ResettableNo