Configures common properties for edge triggering.

An edge trigger occurs when a signal crosses a trigger threshold you specify.


icon

Inputs/Outputs

  • cu16.png trigger coupling

    trigger coupling specifies how to couple the trigger signal.

    Refer to the Trigger Coupling property for defined values.

  • ci32.png trigger slope

    trigger slope specifies either a rising edge or a falling edge to trigger the digitizer.

    Refer to the Trigger Slope property for defined values.

  • civrn.png instrument handle

    instrument handle identifies a particular instrument session.

  • cstr.png trigger source (Channel 0)

    trigger source passes the source you want the digitizer to monitor for a trigger.

    Default Value: Channel 0

    Refer to the Trigger Source property for defined values.

  • cdbl.png trigger level

    trigger level specifies the voltage threshold for the trigger.

    Default Value: 0.0 V

  • cerrcodeclst.png error in

    error in describes error conditions that occur before this node runs. This input provides standard error in functionality.

  • cdbl.png trigger holdoff

    trigger holdoff specifies the length of time the digitizer waits after detecting a trigger before enabling NI-SCOPE to detect another trigger.

    Default Value: 0.0 s

  • cdbl.png trigger delay

    trigger delay specifies how long the digitizer waits after it receives the trigger to start acquiring data.

    Default Value: 0.0 s

  • iivrn.png instrument handle out

    instrument handle out has the same value as the instrument handle.

  • ierrcodeclst.png error out

    error out contains error information. This output provides standard error out functionality.